首页> 外文期刊>Electroanalysis >A Concentric Ring Electrode for a Wall-jet Cell in a Microfluidic Device
【24h】

A Concentric Ring Electrode for a Wall-jet Cell in a Microfluidic Device

机译:用于微流体装置中的壁式喷射电池的同心环电极

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A concentric ring array electrode that amplifies the current signal without redox cycling has been developed for highly sensitive electrochemical detection at a single potential in a microfluidic platform. Herein, the effect of ring-electrode width on the current and current density was examined. A ring-array electrode with widths that decrease from the inner to the outer ring was shown to exhibit the highest sensitivity. This electrode delivered a current density that was approximately 50 % higher than that of a conventionally used disc electrode. We used numerical simulations to further optimize this type of array electrode, which led to a limit of detection for catechol of 6.2 nmol/L. This ring array electrode has great potential for use in a variety of applications because it can be used to detect irreversible targets with a simple apparatus at a single potential and requires no electrode modification to achieve high sensitivity.
机译:在微流体平台中的单一电位下,已经开发了一种同心环形阵列电极,该电极扩增了没有氧化还原循环的电流信号。 这里,检查了环电极宽度对电流和电流密度的影响。 具有从内圈到外圈的宽度的环形阵列电极显示出最高的灵敏度。 该电极输送了比常规使用的盘电极高约50%的电流密度。 我们使用数值模拟来进一步优化这种类型的阵列电极,这导致了6.2 nmol / L的儿茶酚的检测限。 该环形阵列电极具有很大的应用,可以用于各种应用,因为它可以用于在单个电位下用简单的装置检测不可逆的目标,并且不需要电极改性以实现高灵敏度。