The effects of x-ray beam hardening on detective quantum efficiency and radiation dose

机译:X射线束硬化对探测量子效率和辐射剂量的影响

获取原文
获取原文并翻译 | 示例
       

摘要

The goal of this preliminary study was to investigate the effects of x-ray beam hardening on the detective quantum efficiency (DQE) and the radiation dose of an inline x-ray imaging system. The ability to decrease the risk of harmful radiation to the patient without compromising the detection capability would more effectively balance the tradeoff between image quality and radiation dose, and therefore benefit the fields of diagnostic x-ray imaging, especially mammography. The DQE and the average glandular dose were both calculated under the same experimental conditions for a range of beam hardening levels, corresponding to no added beam hardening and two thicknesses each of Rhodium (Rh) and Molybdenum (Mo) filters. The dose calculation results demonstrate a reduction of 15% to 24% for the range of beam hardening levels. The comparison of all quantities comprising the DQE exhibit very close correlation between the results obtained without added beam hardening to the results corresponding to the range of beam hardening levels. For the specific experimental conditions utilized in this preliminary study, the results are an indication that the use of beam hardening holds the potential to reduce the radiation dose without decreasing the performance of the system. Future studies will seek to apply this method in a clinical environment and perform a comprehensive image quality evaluation, in an effort to further evaluate the potential of beam hardening to balance the tradeoff between dose and image quality.
机译:这项初步研究的目的是研究X射线束硬化对在线X射线成像系统的探测量子效率(DQE)和辐射剂量的影响。在不损害检测能力的情况下降低对患者有害辐射的风险的能力将更有效地平衡图像质量和辐射剂量之间的折衷,因此有利于诊断X射线成像(尤其是乳腺摄影)领域。 DQE和平均腺体剂量均在相同的实验条件下针对一定范围的射束硬化水平进行了计算,这相当于没有增加射束硬化,并且铑(Rh)和钼(Mo)滤光片的厚度各为两个。剂量计算结果表明,在射束硬化水平范围内,降低了15%至24%。包含DQE的所有数量的比较显示,在不增加射束硬化的情况下获得的结果与对应于射束硬化级别范围的结果之间具有非常密切的相关性。对于此初步研究中使用的特定实验条件,结果表明,使用束硬化技术有潜力在不降低系统性能的情况下减少辐射剂量。未来的研究将寻求将该方法应用于临床环境并进行全面的图像质量评估,以进一步评估光束硬化的可能性,以平衡剂量和图像质量之间的平衡。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号