首页> 外文期刊>Diabetes technology & therapeutics >The accuracy of home glucose meters in hypoglycemia.
【24h】

The accuracy of home glucose meters in hypoglycemia.

机译:家用血糖仪在低血糖症中的准确性。

获取原文
获取原文并翻译 | 示例
       

摘要

BACKGROUND: Home glucose meters (HGMs) may not be accurate enough to sense hypoglycemia. We evaluated the accuracy and the capillary and venous comparability of five different HGMs (Optium Xceed [Abbott Diabetes Care, Alameda, CA, USA], Contour TS [Bayer Diabetes Care, Basel, Switzerland], Accu-Chek Go [Roche Ltd., Basel, Switzerland], OneTouch Select [Lifescan, Milpitas, CA, USA], and EZ Smart [Tyson Bioresearch Inc., Chu-Nan, Taiwan]) in an adult population. METHODS: The insulin hypoglycemia test was performed to 59 subjects (56 males; 23.6 +/- 3.2 years old). Glucose was measured from forearm venous blood and finger capillary samples both before and after regular insulin (0.1 U/kg) was injected. Venous samples were analyzed in the reference laboratory by the hexokinase method. In vitro tests for method comparison and precision analyses were also performed by spiking the glucose-depleted venous blood. RESULTS: All HGMs failed to sense hypoglycemia to some extend. EZ Smart was significantly inferior in critical error Zone D, and OneTouch Select was significantly inferior in the clinically unimportant error Zone B. Accu-Chek Go, Optium Xceed, and Contour TS had similar performances and were significantly better than the other two HGMs according to error grid analysis or International Organization for Standardization criteria. The in vitro tests were consistent with the above clinical data. The capillary and venous consistencies of Accu-Chek Go and OneTouch Select were better than the other HGMs. CONCLUSIONS: The present results show that not all the HGMs are accurate enough in low blood glucose levels. The patients and the caregivers should be aware of these restrictions of the HGMs and give more credit to the symptoms of hypoglycemia than the values obtained by the HGMs. Finally, these results indicate that there is a need for the revision of the accuracy standards of HGMs in low blood glucose levels.
机译:背景:家用血糖仪(HGM)可能不够准确,无法感知低血糖症。我们评估了五种不同HGM(Optium Xceed [阿伯特糖尿病护理,阿拉米达,加利福尼亚州,美国],Contour TS [拜耳糖尿病护理,巴塞尔,瑞士],Accu-Chek Go [Roche Ltd.,成年人口中的OneTouch Select(美国加利福尼亚州米尔皮塔斯市的Lifescan)和EZ Smart(台湾中南市的Tyson Bioresearch Inc.)。方法:对59名受试者(56名男性; 23.6 +/- 3.2岁)进行了胰岛素低血糖测试。在注射常规胰岛素(0.1 U / kg)之前和之后,从前臂静脉血和手指毛细血管样本中测量葡萄糖。在参考实验室中通过己糖激酶法分析静脉样本。还通过掺入贫葡萄糖的静脉血来进行方法比较和精密度分析的体外测试。结果:所有HGMs均未能在一定程度上感知低血糖。 EZ Smart在严重错误区域D中显着劣势,而OneTouch Select在临床上不重要的错误区域B中显着劣势。错误网格分析或国际标准化组织的标准。体外试验与上述临床数据一致。 Accu-Chek Go和OneTouch Select的毛细管和静脉稠度优于其他HGM。结论:目前的结果表明,并非所有的HGM在低血糖水平上都足够准确。患者和护理人员应了解HGM的这些限制,并且比HGM获得的值更多地归因于低血糖症状。最后,这些结果表明,需要修订低血糖水平的HGM的准确性标准。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号