Accuracy of the StatStrip versus SureStep Flexx glucose meter in neonates at risk of hypoglycemia.

机译:StatStrip与SureStep Flexx血糖仪在患有低血糖风险的新生儿中的准确性。

获取原文
获取原文并翻译 | 示例
       

摘要

The study was performed to evaluate the accuracy of the StatStrip (SS) and SureStep Flexx (SF) glucose meters compared to plasma glucose in infants at risk for neonatal hypoglycemia and to determine the effect of bilirubin and hematocrit on the results. A prospective cross-sectional study was conducted on 172 venous blood glucose samples from infants who had initial low point-of-care (POC) glucose tests measured simultaneously by SS and SF. Plasma glucose levels were compared to both POC instruments, and the effect of bilirubin and hematocrit levels on mean glucose differences were analysed. Mean (SD) plasma glucose was 2.12 (0.45)?mmol/L; (range, 1.11-3.06 mmol/L). Mean (1.96SD) glucose differences of the SS versus SF were 0.21 (0.70) mmol/L and -0.04 (0.78)?mmol/L, respectively. SS sensitivity was 94.7 % with an 86.1 % negative predictive value (NPV) at 2.8 mmol/L, while the SF had a 100 % sensitivity and NPV at the same cut-off level. No correlations were identified between mean glucose differences and either hematocrit or bilirubin levels in both glucose meters. Both the SS and SF glucose meters have limited use when compared to plasma glucose. Hence, they can only be employed as screening tools in at-risk neonates with an appropriate, predetermined cut-off level. Hematocrit and bilirubin levels did not affect the accuracy of both devices.
机译:进行这项研究是为了评估StatStrip(SS)和SureStep Flexx(SF)血糖仪与新生儿血糖不足风险婴儿血浆葡萄糖相比的准确性,并确定胆红素和血细胞比容对结果的影响。前瞻性的横断面研究是对来自婴儿的172份静脉血糖样本进行的,这些样本最初通过SS和SF同时进行了低护理点(POC)葡萄糖测试。将血浆葡萄糖水平与两种POC仪器进行比较,并分析胆红素和血细胞比容水平对平均葡萄糖差异的影响。平均(SD)血浆葡萄糖为2.12(0.45)?mmol / L; (范围1.11-3.06 mmol / L)。 SS与SF的平均(1.96SD)葡萄糖差异分别为0.21(0.70)mmol / L和-0.04(0.78)?mmol / L。 SS灵敏度为94.7%,在2.8 mmol / L时阴性预测值(NPV)为86.1%,而SF在相同的临界值下具有100%的灵敏度和NPV。在两个血糖仪中,平均血糖差异与血细胞比容或胆红素水平之间均未发现相关性。与血浆葡萄糖相比,SS和SF血糖仪的使用都受到限制。因此,它们只能用作具有适当,预定临界值的高危新生儿的筛查工具。血细胞比容和胆红素水平不影响两个设备的准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号