OBJECTIVES: Investigate the interference of Virox(R) disinfectant wipes containing hydrogen peroxide with Lifescan SureStep Flexx(R) glucose meter results. DESIGN AND METHODS: Glucose measurements of control and whole blood samples were performed before and after cleaning with Virox(R) wipes to reveal the kinetics of this interference. RESULTS AND CONCLUSIONS: After a period where the error code 2(b) is given by the device, an overestimation period of glucose results is always present at all levels tested for a duration of hours. As this interference cannot be detected by the operator, hydrogen peroxide containing wipes for sanitizing of the glucose oxidase based glucose meter must be forbidden.
机译:目的:用Lifescan SureStep Flexx血糖仪结果调查含过氧化氢的Virox消毒湿巾的干扰。设计与方法:在用擦拭巾清洁之前和之后,对对照和全血样品进行葡萄糖测量,以揭示这种干扰的动力学。结果与结论:在设备给出错误代码2(b)的一段时间后,在数小时内的所有测试水平上,总会出现对葡萄糖结果的高估期。由于操作员无法检测到这种干扰,因此必须禁止使用含有过氧化氢的抹布对基于葡萄糖氧化酶的血糖仪进行消毒。