【24h】

Design and evaluation of portable shielding case for thyroid ~(131)I monitoring by imaging plate

机译:成像板监测甲状腺〜(131)I便携式屏蔽盒的设计与评估

获取原文
获取原文并翻译 | 示例
           

摘要

A portable shielding case was designed and evaluated for thyroid ~(131)I monitoring by imaging plate (IP) under storage temperature of 0 and 25 C. The background count with shielding case was 20 % of that without shielding case in 25 C after 7 day. The fading rate decreased to 85.3 and 66.4 % after 7 day with and without shielding case. The detection limit was low by shielding and storing at low temperature and the lower detection limits of the IP system are less than the screening level even 7 day after irradiation. The use of shielding case and a low temperature is effective in reducing the fading rate of IP.
机译:设计并设计了便携式屏蔽盒,用于评估成像板(IP)在0和25 C的存储温度下对甲状腺〜(131)I的监测。7以后,带有屏蔽盒的背景计数是25°C时不带屏蔽盒的背景计数的20%天。在有和没有屏蔽的情况下,经过7天的褪色率分别降至85.3和66.4%。由于在低温下屏蔽和保存,检出限较低,即使在照射后7天,IP系统的检出限也低于筛选水平。使用屏蔽盒和低温可有效降低IP的衰落率。