An MRI-compatible semiautomated vacuum-assisted breast biopsy system: Initial feasibility study.

机译:MRI兼容的半自动真空辅助乳房活检系统:初步可行性研究。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To develop an MR-compatible vacuum-assisted core needle breast biopsy system. MATERIALS AND METHODS: A vacuum-assisted breast biopsy system (Mammotome Hand Held; Ethicon Endo-Surgery, USA) was modified for freehand MRI-guided biopsy in an open, interventional 0.5-T scanner (Signa SP; GE, USA). Probes (11 gauge [G]) were fabricated without significant susceptibility artifact. These mate with an electromechanical hand piece and control system that were modified for use within the MRI scanner. A total of 62 breast lesions were simulated in the mammary tissues of six recently lactating sows by injecting between 0.1 and 1.0 mL of an aqueous gel containing dilute gadopentetate dimeglumine (Gd-DTPA) that formed a bright mass on T1-weighted imaging. RESULTS: Mechanical performance was satisfactory. Magnetic susceptibility and radiofrequency (RF) artifacts from the 11-G probe were negligible. T1-weighted fast spin echo (FSE) was used to guide biopsy. Up to eight samples were removed per lesion. Overall, 461 samples were obtained in 493 attempts (94%). Sample weights averaged 54 mg (N = 493) compared to 4.6 mg (N = 24) from 14-G titanium MRI-compatible needles. On average, 59% of the attempted samples yielded target lesion material. CONCLUSION: Preliminary results demonstrate the feasibility of a vacuum-assisted breast biopsy system in the MRI environment. Small 0.1-mL lesions can be biopsied without needle artifacts obscuring the target. J. Magn. Reson. Imaging 2005;21:637-644. (c) 2005 Wiley-Liss, Inc.
机译:目的:开发一种与MR兼容的真空辅助核心针式乳房活检系统。材料和方法:在开放的介入式0.5-T扫描仪(Signa SP; GE,美国)中,对真空辅助的乳房活检系统(Mammotome手持; Ethicon Endo-Surgery,美国)进行了徒手的MRI引导活检。探针(11号[G])制作时没有明显的磁化伪影。它们与机电式机头和控制系统配合使用,该机头和控制系统经过改装,可在MRI扫描仪中使用。通过注入0.1到1.0 mL的含稀释的g丁戊二酸二聚亮氨酸(Gd-DTPA)的水性凝胶在6个最近泌乳的母猪的乳腺组织中模拟了总共62个乳腺病变,在T1加权成像中形成了明亮的肿块。结果:机械性能令人满意。来自11-G探针的磁化率和射频(RF)伪影可以忽略不计。 T1加权快速自旋回波(FSE)用于指导活检。每个病变最多取出八个样本。总体上,在493次尝试中获得了461个样本(94%)。 14-G钛MRI兼容针的平均样品重量为54毫克(N = 493),而4.6毫克(N = 24)。平均而言,尝试的样本中有59%产生了目标病变材料。结论:初步结果证明了真空辅助乳房活检系统在MRI环境中的可行性。可以对0.1 mL的小病变进行活检,而不会出现针状伪影模糊目标的情况。 J.Magn。雷森成像2005; 21:637-644。 (c)2005 Wiley-Liss,Inc.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号