Diagnosing diabetes mellitus: Performance of hemoglobin A1c point-of-care instruments in general practice offices

机译:诊断糖尿病:全科诊所血红蛋白A1c即时医疗器械的性能

获取原文
获取原文并翻译 | 示例
       

摘要

BACKGROUND: Hemoglobin A1c (Hb A1c) measurement by hospital laboratory instruments, but not by pointof-care (POC) instruments, has been recommended for use to diagnose diabetes mellitus. We evaluated results from 13 Hb A1c external quality assurance (EQA) surveys over a 6-year period in Norway, from both POC instruments used in general practice (GP) offices and instruments in hospital laboratories, against the analytical quality specifications recommended for use of Hb A1c to diagnose diabetes mellitus. METHODS: All GP offices (n = 1288) and hospital laboratories (n = 52) measuring Hb A1c in Norway participated in the EQA survey. The percentage of participants that performed measurements within the quality specifications was calculated. Pooled within-laboratory CVs were estimated for the Afinion, DCA 2000, DCA 2000 +, DCA VantageTM, and Nycocard Hb A1c Reader instruments and for hospital laboratory instruments. RESULTS: Between 60% to 90% of Afinion and DCA users and hospital laboratories performed Hb A1c measurements within the quality specifications for both trueness (6.0%) and imprecision (CV ≤ 2.0%) at 2 levels in each EQA survey. The pooled within-laboratory CVs for the Afinion andDCAinstruments and hospital laboratories were below the recommended limit of 2.0% for most of the surveys. CONCLUSIONS: A large proportion of GP offices using Afinion and DCA POC instruments to measure Hb A1c fulfill the analytical quality specifications for diagnosing diabetes mellitus, and these instruments demonstrate analytical quality comparable to that of hospital laboratory instruments. When GP offices participate in a stringent quality assurance program and generate Hb A1c measurements that meet analytical quality specifications, these measurements can be recommended for use to diagnose diabetes mellitus.
机译:背景:已建议通过医院实验室仪器而不是现场护理(POC)仪器进行血红蛋白A1c(Hb A1c)测量,以诊断糖尿病。我们评估了挪威在6年内进行的13次Hb A1c外部质量保证(EQA)调查得出的结果,这些结果均来自推荐用于实验室的普通质量(GP)办公室使用的POC仪器和医院实验室的仪器, Hb A1c诊断糖尿病。方法:挪威所有的全科医生办公室(n = 1288)和医院实验室(n = 52)都测量了Hb A1c。计算在质量规格范围内进行测量的参与者的百分比。评估了Afinion,DCA 2000,DCA 2000 +,DCA VantageTM和Nycocard Hb A1c Reader仪器以及医院实验室仪器的实验室内部CV汇总。结果:在每次EQA调查中,真实性(6.0%)和不精确性(CV≤2.0%)的质量指标中,有60%至90%的Afinion和DCA用户和医院实验室进行了Hb A1c测量。对于大多数调查,Afinion和DCA仪器以及医院实验室的实验室内CV汇总值均低于建议的2.0%的限制。结论:很大一部分使用Afinion和DCA POC仪器测量Hb A1c的全科医生办公室都满足诊断糖尿病的分析质量要求,并且这些仪器的分析质量可与医院实验室仪器相媲美。当GP办事处参加严格的质量保证计划并生成符合分析质量规格的Hb A1c测量值时,可以建议将这些测量值用于诊断糖尿病。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号