On-line coupling of solid-phase extraction with mass spectrometry for the analysis of biological samples III. Determination of prednisolone in serum

机译:固相萃取与质谱的在线耦合,用于分析生物样品III。血清泼尼松龙的测定

获取原文
获取原文并翻译 | 示例
       

摘要

Solid-phase extraction (SPE) was directly coupled to mass spectrometry (MS) to assess the feasibility of the system for the rapid determination of prednisolone in serum. A C_(18) stationary phase allowed washing of the cartridge with 25% methanol. Elution was performed by switching the methanol percentage from 25% in the washing step to 50% during elution. The high flow-rates during the extraction (5.0 ml/min) combined with ion-trap MS detection resulted in a total analysis time of 4 min. Some tailing of the prednisolone peak was observed. However, the tailing was found acceptable, since by this elution procedure most matrix compounds were preventd from eluting from the cartridge. Some matrix interference was still observed with a triple-quadrupole MS, even in the multiple reaction monitoring mode. This resulted in a detection limit (LOD) of about 10 ng/ml. The matrix interference and the LOD were similar for atmospheric pressure chemical ionisation and atmospheric pressure photo ionisation. Applying an ion-trap MS in the MS-MS mode resulted in cleaner chromatograms. Due to extensive fragmentation of prednisolone, the LOD was not lower than about 5 ng/ml prednisolone in serum, and a limit of quantitation of about 10 ng/ml (relative standard deviation <15%) was observed.
机译:固相萃取(SPE)直接与质谱(MS)耦合以评估该系统快速测定血清泼尼松龙的可行性。 C_(18)固定相允许用25%的甲醇洗涤小柱。通过将甲醇百分比从洗涤步骤中的25%切换到洗脱期间的50%进行洗脱。萃取过程中的高流速(5.0 ml / min)与离子阱MS检测相结合,导致总分析时间为4分钟。观察到泼尼松龙峰有些拖尾。但是,发现拖尾是可以接受的,因为通过该洗脱程序,大多数基质化合物都无法从药筒中洗脱出来。即使在多反应监测模式下,使用三重四极杆质谱仪仍观察到一些基质干扰。这导致约10 ng / ml的检测限(LOD)。对于大气压化学电离和大气压光电离,基质干扰和LOD相似。在MS-MS模式下应用离子阱质谱仪可得到更干净的色谱图。由于泼尼松龙的广泛片段化,血清中的LOD不低于约5 ng / ml泼尼松龙,并且观察到的定量限为约10 ng / ml(相对标准偏差<15%)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号