首页> 外文期刊>The Journal of Bone and Joint Surgery. American Volume >Intrinsic constraint of unlinked total elbow replacements--the ulnotrochlear joint.
【24h】

Intrinsic constraint of unlinked total elbow replacements--the ulnotrochlear joint.

机译:不相关的全肘关节置换的固有约束-尺骨关节。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Many unlinked total elbow replacement designs with radically differing articular geometries exist, suggesting that there is no consensus regarding an optimal design. A feature inherent to the articular design is the intrinsic constraint afforded to the joint by the implant. Our aim was to compare the intrinsic constraints of unlinked implants with that of the normal ulnotrochlear joint. METHODS: We tested twelve cadaveric ulnotrochlear joints with a custom-made multiple-axis materials testing machine. With compressive loads ranging from 10 to 100 N, the joints were moved in either valgus or varus directions at 90 degrees of flexion. The ulnotrochlear components from a single example of five medium-sized unlinked elbow replacements (Ewald, Kudo, Pritchard ERS, Sorbie-Questor, and Souter-Strathclyde) were also tested. The recorded measurements included the torques and forces, angular displacement, and axial displacement of the humerus relative to the ulna. RESULTS: In general, the peak torqueand the constraint ratio significantly increased with increasing compressive load for the implants as well as for the normal elbow. In valgus displacement, the Souter-Strathclyde implant had the highest and the Sorbie-Questor had the smallest peak torque and the Souter-Strathclyde had the highest and the Ewald had the smallest constraint ratio. In varus displacement, the Kudo had the highest and the Ewald had the smallest peak torque and constraint ratio. CONCLUSIONS: The constraint ratio is a characteristic that is useful for describing elbow joint behavior and for comparing the behavior of implants with that of the human elbow. Of the unlinked implants tested, the Souter-Strathclyde and Kudo prostheses most closely approximated the behavior of the human elbow joint. Implants that resemble the human elbow in appearance do not replicate normal behavior consistently, whereas other implants that do not resemble the human elbow closely do not deviate markedly from human behavior. Thus, much basic information about elbow form and function is needed to improve the performance of total elbow prostheses.
机译:背景:存在许多具有根本不同的关节几何形状的未链接的全肘关节置换设计,这表明关于最佳设计尚无共识。关节设计固有的特征是植入物对关节的固有约束。我们的目的是比较未连接的植入物与正常尺骨关节的固有约束。方法:我们使用定制的多轴材料测试机测试了十二个尸体尺骨关节。在10到100 N的压缩载荷下,关节以90度屈曲向外翻或内翻方向移动。还测试了来自五个中型未链接肘替代物(Ewald,Kudo,Pritchard ERS,Sorbie-Questor和Souter-Strathclyde)的单个示例的ulnotrochlear组件。记录的测量值包括肱骨相对于尺骨的扭矩和力,角位移和轴向位移。结果:总体上,随着种植体和正常肘部压缩载荷的增加,峰值扭矩和约束比也显着增加。在外翻位移中,Souter-Strathclyde植入物的峰值扭矩最高,而Sorbie-Questor的峰值扭矩最小,Suterbie-Strathclyde的植入物峰值最高,而Ewald的约束比最小。在内翻位移中,Kudo最高,Ewald的峰值扭矩和约束比最小。结论:约束比是一种特性,可用于描述肘关节的行为以及将植入物的行为与人类肘部的行为进行比较。在测试的未连接的植入物中,Souter-Strathclyde和Kudo假体最接近人类肘关节的行为。在外观上类似于人类肘部的植入物不能始终如一地复制正常行为,而其他与人类肘部不相似的植入物则不会明显偏离人类行为。因此,需要许多有关肘部形状和功能的基本信息,以改善整个肘部假体的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号