Study of High-Resolution Mass Spectrometry Technology as a Replacement for Tandem Mass Spectrometry in the Field of Quantitative Pesticide Residue Analysis

机译:农药残留定量分析领域中代替串联质谱的高分辨率质谱技术研究

获取原文
获取原文并翻译 | 示例

摘要

A validated LC/MS/MS-based multiresidue pesticide method was converted to an LC high-resolution MS single-stage Orbitrap platform. No changes regarding the cleanup and LC were made. Optimization of high-resolution MS-specific parameters and interface settings was kept at a minimum. The aim was to explore the capability of current Orbitrap technology to substitute for LC/MS/MS technology. The test included the quantitative performance (sensitivity, selectivity, linearity, accuracy, and precision) of some 240 analytes in three different matrixes. The LC/MS/MS instrumentation was operated at the edge of its technical limitations. A further extension of the number of analytes for LC/MS/MS would require the use of even narrower dwell times, significantly reducing sensitivity and reproducibility of measurement. No such limitations exist for the high-resolution technology. Similar performance was observed for both technologies. A current drawback of the high-resolution technology is the speed of data processing, which took significantly longer than for LC/MS/MS data due to the limited capabilities of the software.
机译:经过验证的基于LC / MS / MS的多残留农药方法已转换为LC高分辨率MS单级Orbitrap平台。没有关于清理和LC的更改。高分辨率MS特定参数和接口设置的优化保持在最低水平。目的是探索当前Orbitrap技术替代LC / MS / MS技术的能力。该测试包括三种不同基质中约240种分析物的定量性能(灵敏度,选择性,线性,准确性和精密度)。 LC / MS / MS仪器在其技术局限性范围内运行。 LC / MS / MS的分析物数量的进一步扩展将需要使用更窄的停留时间,从而大大降低了测量的灵敏度和可重复性。高分辨率技术不存在此类限制。两种技术的性能相似。高分辨率技术的当前缺点是数据处理速度,由于该软件的功能有限,它花费的时间明显长于LC / MS / MS数据。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号