首页> 外文期刊>Journal of AOAC International >A comparative study of first-derivative spectrophotometry and column high-performance liquid chromatography applied to the determination of repaglinide in tablets and for dissolution testing
【24h】

A comparative study of first-derivative spectrophotometry and column high-performance liquid chromatography applied to the determination of repaglinide in tablets and for dissolution testing

机译:一阶导数分光光度法与色谱柱高效液相色谱法在片剂中瑞格列奈测定和溶出度测试中的比较研究

获取原文
获取原文并翻译 | 示例
           

摘要

A fast and reliable method for the determination of repaglinide is highly desirable to support formulation screening and quality control. A first-derivative UV spectroscopic method was developed for the determination of repaglinide in tablet dosage form and for dissolution testing. First-derivative UV absorbance was measured at 253 nm. The developed method was validated for linearity, accuracy, precision, limit of detection (LOD), and limit of quantitation (LOQ) in comparison to the U.S. Pharmacopeia (USP) column high-performance liquid chromatographic (HPLC) method. The first-derivative UV spectrophotometric method showed excellent linearity [correlation coefficient (r) = 0.9999] in the concentration range of 1-35 mu g/mL and precision (relative standard deviation < 1.5%). The LOD and LOQ were 0.23 and 0.72 mu g/mL, respectively, and good recoveries were achieved (98-101.8%). Statistical comparison of results of the first-derivative UV spectrophotometric and the USP HPLC methods using the t-test showed that there was no significant difference between the 2 methods. Additionally, the method was successfully used for the dissolution test of repaglinide and was found to be reliable, simple, fast, and inexpensive.
机译:为了支持制剂筛选和质量控制,非常需要一种快速可靠的瑞格列奈测定方法。开发了一种一阶导数紫外光谱法,用于测定片剂剂型中的瑞格列奈和进行溶出度测试。一阶导数UV吸光度在253 nm下测量。与美国药典(USP)色谱柱高效液相色谱(HPLC)方法相比,该开发方法在线性,准确性,精密度,检测限(LOD)和定量限(LOQ)方面得到了验证。一阶导数紫外分光光度法在1-35μg / mL的浓度范围内显示出极好的线性[相关系数(r)= 0.9999]和精确度(相对标准偏差<1.5%)。 LOD和LOQ分别为0.23和0.72μg / mL,并获得了良好的回收率(98-101.8%)。使用t检验的一阶导数紫外分光光度法和USP HPLC方法的结果的统计比较表明,两种方法之间没有显着差异。此外,该方法已成功用于瑞格列奈的溶出度测试,并且发现该方法可靠,简单,快速且便宜。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号