OBJECTIVE: To assess the implications of falsely elevated glucose readings measured with glucose dehydrogenase pyrroloquinolinequinone (GDH-PQQ) test strips. RESEARCH DESIGN AND METHODS: We conducted a review of the Food and Drug Administration's Manufacturer and User Facility Device Experience database and medical literature for adverse events (AEs) associated with falsely elevated glucose readings with GDH-PQQ test strips in the presence of interfering sugars. RESULTS: Eighty-two reports were identified: 16 (20%) were associated with death, 46 (56%) with severe hypoglycemia, and 12 (15%) with nonsevere hypoglycemia. In eight reports (10%), the AE was not described. Forty-two events (51%) occurred in the U.S. Although most events occurred in hospitalized patients, at least 14 (17%) occurred in outpatients. Agents most commonly associated with AEs were icodextrin-containing peritoneal dialysate and maltose-containing intravenous immune globulin. CONCLUSIONS: GDH-PQQ test strips pose a safety risk to insulin-using patients treated with agents containing or metabolized to interfering sugars.
机译:目的:评估使用葡萄糖脱氢酶吡咯并喹啉醌(GDH-PQQ)试纸测得的葡萄糖读数错误升高的含义。研究设计和方法:我们审查了食品和药物管理局的制造商和用户设施设备体验数据库,以及在存在干扰糖的情况下,使用GDH-PQQ试纸与葡萄糖读数错误升高相关的不良事件(AE)的医学文献。结果:八十二报告被确定:与死亡相关的16(20%),严重低血糖的46(56%),和非严重低血糖的12(15%)。在八份报告中(10%),没有描述不良事件。在美国发生了42起事件(51%)。尽管大多数事件发生在住院患者中,但在门诊患者中至少发生了14起(17%)。最常与AE相关的药物是含艾考糊精的腹膜透析液和含麦芽糖的静脉内免疫球蛋白。结论:GDH-PQQ试纸对使用含糖或被代谢成干扰糖的药物治疗的使用胰岛素的患者构成安全风险。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号