首页> 外文期刊>High Energy Chemistry >Comparison between the distributions of excitation temperatures and the analytical line intensities of noble metals in two-jet arc plasmatrons used in atomic emission analysis
【24h】

Comparison between the distributions of excitation temperatures and the analytical line intensities of noble metals in two-jet arc plasmatrons used in atomic emission analysis

机译:原子发射分析中使用的两喷弧等离子加速器的激发温度分布与贵金属分析线强度的比较

获取原文
获取原文并翻译 | 示例
           

摘要

The results of a study of DGP-50 and new EDP-355 argon two-jet plasmatrons, which are used in the atomic emission analysis of geological samples, are reported. The distributions of temperatures and the analytical line intensities of noble metals with torch height were measured for either of the plasmatrons. It was found that the temperature reached in the EDP-355 plasmatron was higher than that in the DGP-50 plasmatron at equal flow rates of a plasma-forming gas. A decrease in the flow rate of a plasma-forming gas in the EDP-355 plasmatron resulted in an increase in the analytical lines of noble metals by a factor of 1.5-2. It was found experimentally that the EDP-355 plasmatron is a promising source for emission spectral analysis in the determination of noble metals.
机译:报道了用于地质样品原子发射分析的DGP-50和新型EDP-355氩气两喷嘴等离子体加速器的研究结果。测量了两个等离子加速器的温度分布和贵金属与焊炬高度的分析线强度。发现在等离子形成气体的相同流速下,EDP-355等离子加速器中达到的温度高于DGP-50等离子加速器中的温度。 EDP​​-355等离子加速器中等离子形成气体流速的降低导致贵金属的分析线增加了1.5-2倍。实验发现,EDP-355等离子加速器是用于确定贵金属的发射光谱分析的有希望的来源。