VALIDATION OF AN INDEXED RADIOTHERAPY HEAD POSITIONING DEVICE FOR USE IN DOGS AND CATS

机译:用于狗和猫的独立放射治疗头部定位设备的验证

获取原文
获取原文并翻译 | 示例
           

摘要

Setup variability affects the appropriate delivery of radiation and informs the setup margin required to treat radiation patients. Twenty-four veterinary patients with head and neck cancers were enrolled in this prospective, cross-sectional study to determine the accuracy of an indexed board immobilization device for positioning. Couch position values were defined at the first treatment based on setup films. At subsequent treatments, patients were moved to the previously defined couch location, orthogonal films were acquired, table position was modified, and displacement was recorded. The mean systematic displacement, random displacement, overall displacement, and mean displacement values of the three-dimensional (3D) vector were calculated. Three hundred thirty-two pairs of orthogonal setup films were analyzed for displacement in cranial-caudal, lateral, and dorsal-ventral directions. The mean systematic displacements were 0.5, 0.8, and 0.5 mm, respectively. The mean random displacements were 1.0, 1.1, and 0.7 mm, respectively. The overall displacements were 1.1, 1.4, and 0.9 mm, respectively. The mean 3D vector value was 1.6 mm with a standard deviation of 1.2 mm. Ninety-five percent of the vectors were <3.6 mm. These values were compared to data obtained with a previously used immobilization device. A t-test was used to compare the two devices. The 3D vector, random displacement in all directions, and overall displacement in the cranial-caudal and dorsal-ventral directions were significantly smaller than displacements with the previous device. The precision and accuracy of the indexed board device was superior to the historical head and neck device.
机译:设置的可变性会影响放射线的适当传送,并告知治疗放射线患者所需的设置余量。这项前瞻性,横断面研究招募了二十四位患有头颈癌的兽医患者,以确定用于定位的分度板固定装置的准确性。沙发位置值是在第一次治疗时根据胶片设定的。在随后的治疗中,将患者转移到先前定义的卧榻位置,获取正交胶片,修改工作台位置,并记录位移。计算了三维(3D)向量的平均系统位移,随机位移,总位移和平均位移值。分析了三百二十三对正交设置膜在颅-尾,横向和背-腹方向的位移。平均系统位移分别为0.5、0.8和0.5 mm。平均随机位移分别为1.0、1.1和0.7毫米。总位移分别为1.1、1.4和0.9毫米。 3D矢量的平均值为1.6毫米,标准偏差为1.2毫米。百分之九十五的载体<3.6mm。将这些值与先前使用的固定装置获得的数据进行比较。使用t检验比较这两种设备。 3D向量,在所有方向上的随机位移以及在颅-尾和背-腹方向上的总体位移都比以前的设备的位移小得多。分度板装置的精度和准确性优于历史上的头颈装置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号