首页> 外文期刊>Journal of applied clinical medical physics / >Investigation on the performance of dedicated radiotherapy positioning devices for MR scanning for prostate planning
【24h】

Investigation on the performance of dedicated radiotherapy positioning devices for MR scanning for prostate planning

机译:MR扫描前列腺计划专用放射治疗定位设备性能的研究

获取原文
           

摘要

The purpose of this study was to investigate performance of the couch and coil mounts designed for MR-simulation prostate scanning using data from ten volunteers. Volunteers were scanned using the standard MR scanning protocol with the MR coil directly strapped on the external body and the volunteer lying on the original scanner table. They also were scanned using a MR-simulation table top and pelvic coil mounts. MR images from both setups were compared in terms of body contour variation and image quality effects within particular organs of interest. Six-field conformal plans were generated on the two images with assigned bulk density for dose calculation. With the MR-simulation devices, the anterior skin deformation was reduced by up to 1.7 cm. The hard tabletop minimizes the posterior body deformation which can be up to 2.3 cm on the standard table, depending on the weight of volunteer. The image signal-to-noise ratio reduced by 14% and 25% on large field of view (FOV) and small FOV images, respectively, after using the coil mount; the prostate volume contoured on two images showed difference of 1.05 ± 0.66 cm 3 . The external body deformation caused a mean dose reduction of 0.6 ± 0.3 Gy , while the coverage reduced by 22 % ± 13 % and 27 % ± 6 % in V 98 and V 100 , respectively. A dedicated MR simulation setup for prostate radiotherapy is essential to ensure the agreement between planning anatomy and treatment anatomy. The image signal was reduced after applying the coil mount, but no significant effect was found on prostate contouring.PACS numbers: 87.55.D-, 87.61.-c, 87.57.C-
机译:这项研究的目的是使用来自十个志愿者的数据来研究用于MR模拟前列腺扫描的沙发和线圈架的性能。使用标准的MR扫描协议对志愿者进行扫描,将MR线圈直接绑在体外,而志愿者则躺在原始扫描仪桌上。还使用MR模拟桌面和骨盆线圈架对其进行了扫描。比较了两种设置的MR图像,这些图像在特定感兴趣器官内的身体轮廓变化和图像质量影响方面进行了比较。在两个图像上生成六场共形图,并分配了堆积密度以进行剂量计算。使用MR模拟设备,前皮肤变形最多可减少1.7 cm。坚硬的桌面可最大程度地减少后方身体变形,根据志愿者的体重,后方身体变形在标准桌子上最大可达2.3厘米。使用线圈架后,在大视场(FOV)和小视场图像上,图像信噪比分别降低了14%和25%。在两幅图像上勾勒出的前列腺体积显示相差1.05±0.66 cm 3。体外变形导致平均剂量降低0.6±0.3 Gy,而V 98和V 100的覆盖率分别降低22%±13%和27%±6%。前列腺放射疗法专用的MR模拟设置对于确保计划解剖结构与治疗解剖结构之间的一致性至关重要。施加线圈支架后图像信号减弱,但对前列腺轮廓没有发现明显影响。PACS编号:87.55.D-,87.61.-c,87.57.C-

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号