首页> 外文期刊>Tribology International >Characterisation of the oxide film on the taper interface from retrieved large diameter metal on polymer modular total hip replacements
【24h】

Characterisation of the oxide film on the taper interface from retrieved large diameter metal on polymer modular total hip replacements

机译:从聚合物模块全髋关节置换物中取回的大直径金属表征锥形界面上的氧化膜

获取原文
获取原文并翻译 | 示例
           

摘要

Many metal joint failures have been associated with adverse local tissue reactions due to the response of the body to wear debris and corrosion products, released from the bearing surfaces or the taper interfaces. The oxide film on the CoCrMo plays an important role in the electrochemical behaviour, however, there is a lack of quantitative data on the structure and distribution of oxides following in vivo operation. In the present study, detailed analysis of the surface layer on the taper interfaces is provided. Two retrieved taper interfaces were analysed. Site-specific FIB/TEM cross-sections show evidence of the oxide film and the carbonaceous layer. High resolution TEM results show the oxide film has porous texture and EELS confirmed it is chromium oxide. (C) 2014 Elsevier Ltd. All rights reserved.
机译:由于身体对从轴承表面或锥度界面释放的磨损碎片和腐蚀产物的反应,许多金属接头失效都与不良的局部组织反应有关。 CoCrMo上的氧化膜在电化学行为中起着重要作用,但是,缺乏体内操作后氧化物的结构和分布的定量数据。在本研究中,提供了锥度界面上的表面层的详细分析。分析了两个检索到的锥度接口。特定位置的FIB / TEM横截面显示出氧化膜和碳质层的迹象。高分辨率TEM结果表明,氧化膜具有多孔质感,EELS证实其为氧化铬。 (C)2014 Elsevier Ltd.保留所有权利。