This study reports the results and early failures using the Graft Augmentation Prosthesis ring in the reconstruction of acetabular defects encountered during total hip arthroplasty. Seventeen consecutive Graft Augmentation Prosthesis rings were used during 7 complex primary and 10 revision hip arthroplasties. Five patients died during the follow-up period. Of the remaining 12 patients, 7 had been revised at an average of 5 years follow-up. Five cases were revised because of fatigue failure of the implant associated with allograft resorption. Two cases were revised for recurrent dislocations. Because of this high mechanical failure rate (5 of 12 cases at only 5 years follow-up), we have abandoned this device in favor of implants with more mechanical strength.
机译:这项研究报告了在全髋关节置换术中遇到的髋臼缺损重建中使用嫁接增强假体环的结果和早期失败。在7例复杂的原发性和10例修订的髋关节置换术中使用了十七个连续的Graft Augmentation假体环。在随访期间有五名患者死亡。其余12例患者中,有7例平均随访5年。由于同种异体移植物吸收相关的植入物疲劳衰竭,对五例进行了修订。针对复发性脱位的两个案例进行了修订。由于如此高的机械故障率(仅在5年的随访中12例中有5例),我们放弃了该设备,转而使用具有更高机械强度的植入物。