首页> 外文期刊>The Journal of arthroplasty >Isocentric 3-Dimensional C-Arm Imaging of Component Alignments in Total Knee Arthroplasty With Potential Intraoperative and Postoperative Applications
【24h】

Isocentric 3-Dimensional C-Arm Imaging of Component Alignments in Total Knee Arthroplasty With Potential Intraoperative and Postoperative Applications

机译:等距3维C臂成像的全膝关节置换术中潜在的术中和术后应用中的组件对齐方式。

获取原文
获取原文并翻译 | 示例
       

摘要

An intraoperative imaging tool for total knee arthroplasty could help avoid poor clinical outcomes related to malalignment. We investigated the feasibility of using isocentric (ISO-C) fluoroscopic imaging for this purpose. Three-dimensional ISO-C and computed tomographic (CT) images were acquired from 6 cadaveric specimens implanted with standard knee arthroplasty components and analyzed to determine rotational alignments. In comparison with standard CT measures, the ISO-C-based measures had overall accuracies (determined as root mean square error) of 0.8?? and 1.3?? and corresponding SDs of 1.3?? and 1.4?? for the femoral and tibial components, respectively. With ISO-C imaging, it is possible to measure rotational alignment in knee arthroplasty with accuracy and repeatability comparable with CT. Isocentric imaging has strong potential as an intraoperative tool to accurately align arthroplasty components. ? 2013 Elsevier Inc.
机译:术中全膝关节置换术的成像工具可以帮助避免与错位相关的不良临床结果。我们研究了为此目的使用等心(ISO-C)荧光透视成像的可行性。从植入了标准膝关节置换组件的6个尸体标本中获取三维ISO-C和计算机断层扫描(CT)图像,并进行分析以确定旋转方向。与标准的CT测量相比,基于ISO-C的测量的总体精度(确定为均方根误差)为0.8?。和1.3 ??和相应的SD为1.3?和1.4 ??分别用于股骨和胫骨组件。通过ISO-C成像,可以以与CT相当的准确性和可重复性来测量膝关节置换术中的旋转对准。等轴心成像具有作为术中工具准确对准关节成形术组件的强大潜力。 ? 2013爱思唯尔公司

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号