BACKGROUND AND AIM OF THE STUDY: A 'Hemispherical' model of aortic valve geometry was developed in which normal human cadaveric aortic valve leaflets could be represented as three hemispheres nested within a cylindrical aorta. By mathematically describing the junction between the leaflets and aorta, the normal three-dimensional annular geometry of the aortic valve could be defined. In this study, a prototype annuloplasty frame based on this model was tested as a repair device in isolated porcine aortic root preparations. METHODS: Eight isolated porcine aortic roots were perfused with water at a constant pressure of 75 mmHg, and valve leakage was measured before and after the vertical incision of one to three sub-commissural areas to create valve incompetence. Annuloplasty frames then were sutured directly to the valve annuli, and the degree of leakage re-determined. The model predicted that the leaflet free-edge length could be used to define the valve diameter necessary to create competence, and this interaction was tested by varying frame size within the protocol. Post-repair leaflet function also was evaluated using high-speed video. Differences were assessed using a two-tailed paired t-test. RESULTS: All eight normal porcine aortic valves had negligible baseline leak. With sub-commissural incision, the mean (+/- SD) regurgitant volume was increased to 522 +/- 378 ml/min. After annuloplasty frame insertion, the measured leak decreased to 52 +/- 40 ml/min (p = 0.004). The best results were obtained with a frame diameter 5-6 mm smaller than the leaflet free-edge length. Data recorded using high-speed video suggested a normal dynamic valve function after frame insertion. CONCLUSION: A three-dimensional 'Hemispherical' intra-annular annuloplasty frame was developed for aortic valve repair. Insertion of the frame into the disrupted porcine valve annuli routinely and effectively restored valve competence. These positive initial results support the continued testing and application of this aortic annuloplasty device.
机译:研究的背景和目的:建立了一个“半球”主动脉瓣几何模型,其中正常人的尸体主动脉瓣小叶可以表示为嵌套在圆柱状主动脉内的三个半球。通过数学描述小叶和主动脉之间的连接,可以定义主动脉瓣的正常三维环状几何形状。在这项研究中,基于此模型的原型瓣环成形术框架已作为隔离猪主动脉根制剂的修复装置进行了测试。方法:在75 mmHg的恒定压力下,对8个离体的猪主动脉根进行灌注,并在垂直切开1至3个子接合区以产生瓣膜功能不全之前和之后测量瓣膜渗漏。然后将瓣环成形术框架直接缝合至瓣膜瓣环,并重新确定渗漏程度。该模型预测,小叶的自由边缘长度可用于定义创造能力所必需的瓣膜直径,并且通过在协议中更改框架大小来测试这种相互作用。还使用高速视频评估了维修后的传单功能。使用两尾配对t检验评估差异。结果:所有八个正常的猪主动脉瓣基线漏泄均可以忽略不计。使用亚合缝切口时,平均(+/- SD)反流体积增加到522 +/- 378 ml / min。插入瓣环成形术框架后,测得的泄漏降低至52 +/- 40 ml / min(p = 0.004)。框架直径比小叶的自由边缘长度小5-6 mm,可获得最佳结果。使用高速视频记录的数据表明,插入框架后,动态阀功能正常。结论:开发了三维“半球形”瓣环成形术框架,用于主动脉瓣修复。常规地将框架插入破裂的猪瓣膜环中并有效地恢复了瓣膜功能。这些积极的初步结果支持了这种主动脉瓣环成形术设备的继续测试和应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号