The aim of this work was to study the influence of patient thickness, tube voltage and image detector on patient dose, contrast and ideal observer signal-to-noise ratio (SNR_I), for pathological details positioned at different regions in the image in posterior—anterior (PA) chest radiology. A Monte Carlo computational model was used to compute measures of physical image quality (contrast, SNR_I) and patient effective dose, E. Two metastasis-like details positioned in the central right lung and right lung near the spine, respectively, were studied. The tube voltage was varied between 100 and 150 kV and the patient thickness between 20 and 28 cm. Both, a computed radiography (CR) system and a direct radiography (DR) system, were investigated. The DR system provides both lower doses and better image quality compared with the CR system. The SNR_I~2/E is ~2.9 times higher for the DR system compared with the CR system.
机译:这项工作的目的是研究患者厚度,管电压和图像检测器对患者剂量,对比度和理想观察者信噪比(SNR_I)的影响,以了解位于后方图像中不同区域的病理学细节,前(PA)胸部放射学。使用蒙特卡洛计算模型来计算物理图像质量(对比度,SNR_I)和患者有效剂量E的量度。研究了分别位于脊柱附近右中肺和右肺的两个类似转移的细节。管电压在100 kV至150 kV之间变化,患者厚度在20 cm至28 cm之间。研究了计算机射线照相(CR)系统和直接射线照相(DR)系统。与CR系统相比,DR系统可提供更低的剂量和更好的图像质量。与CR系统相比,DR系统的SNR_I〜2 / E高约2.9倍。