Quantifying computed radiography (CR) and digital radiography (DR) image quality and patient dose for pediatric radiology.

机译:量化计算机放射线照相(CR)和数字放射线照相(DR)的图像质量以及儿科放射学的患​​者剂量。

获取原文
获取原文并翻译 | 示例

摘要

With the advent of digital imaging modalities, there is a potential for dose reduction especially for pediatric patients due to the wide dynamic range and the image processing capabilities of digital systems. A Canon digital radiography (DR) system and an Agfa computed radiography (CR) system are currently in use at Shands Hospital at the University of Florida. Using objective measures of image quality (modulation transfer function (MTF), noise power spectrum (NPS), and integral detective quantum efficiency (IDQE)), these systems can be evaluated and compared for pediatric chest imaging. Using MIRD tissue-equivalent pediatric phantoms constructed at the University of Florida, organ absorbed and effective doses of chest examinations can be measured. With these measurements, a relationship between image quality and dose is used to predict appropriate techniques used for pediatric chest imaging.;The radiographic techniques developed in this research for the Agfa CR system were identical to the radiographic techniques that a technologist would select for a patient the same size and weight of the pediatric phantoms. The method for predicting appropriate radiographic techniques is useful for setting up new systems such as the Canon DR system at Shands at the University of Florida which is not used on patients under 4 years of age. This research determined that the Canon DR system had superior image quality, as measured by the HDQE, at lower doses than the Agfa CR system at standard resolution.
机译:随着数字成像模式的出现,由于数字系统的宽动态范围和图像处理能力,特别是对于小儿患者而言,存在降低剂量的潜力。佛罗里达大学的Shands医院目前正在使用佳能数字放射(DR)系统和Agfa计算机放射(CR)系统。使用客观的图像质量度量(调制传递函数(MTF),噪声功率谱(NPS)和积分检测量子效率(IDQE)),可以对这些系统进行评估并进行儿科胸部成像的比较。使用佛罗里达大学建造的MIRD组织等效的儿科幻象,可以测量器官吸收和胸部检查的有效剂量。通过这些测量,可以使用图像质量和剂量之间的关系来预测用于儿科胸部成像的适当技术。该研究为Agfa CR系统开发的放射线照相技术与技术人员为患者选择的放射线照相技术相同尺寸和重量相同的儿科幻影。适当的射线照相技术的预测方法对于建立新系统非常有用,例如佛罗里达大学Shands的Canon DR系统,该系统不适用于4岁以下的患者。这项研究确定,通过HDQE测量,佳能DR系统在标准分辨率下的剂量要低于爱克发CR系统,其图像质量要好于HDQE。

著录项

  • 作者

    Johnson, Kennita A.;

  • 作者单位

    University of Florida.;

  • 授予单位 University of Florida.;
  • 学科 Biomedical engineering.;Medical imaging.
  • 学位 Ph.D.
  • 年度 2003
  • 页码 267 p.
  • 总页数 267
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

  • 入库时间 2022-08-17 11:44:48

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号