Digital Tomosynthesis of the Chest: Comparison of Patient Exposure Dose and Image Quality between Standard Default Setting and Low Dose Setting

机译:胸部数字断层合成:标准默认设置和低剂量设置之间患者暴露剂量和图像质量的比较

获取原文
           

摘要

Objective To determine the optimum low dose (LD) digital tomosynthesis (DT) setting, and to compared the image quality of the LD DT with that of the standard default (SD) DT. Materials and Methods Nine DT settings, by changing tube voltage, copper filter, and dose ratio, were performed for determining the LD setting. Among combinations of DT setting, a condition providing the lowest radiation dose was determined. Eighty artificial nodules less than 1 cm in diameter (subcentimeter nodules: 40, micronodules less than 4 mm: 40) were attached to a Styrofoam and a diaphragm of the phantom. Among these, 38 nodules were located at the periphery of the lung (thin area) and 42 nodules were located at the paravertebral or sub-diaphragmatic area (thick area). Four observers counted the number of nodules detected in the thick and thin areas. The detection sensitivity in SD and LD settings were calculated separately. Data were analyzed statistically. Results The lowest LD setting was a combination of 100 kVp, 0.3 mm additional copper filter, and a 1 : 5 dose ratio. The effective dose for the LD and SD settings were 62 μSv and 140 μSv, separately. A 56.7% dose reduction was achieved in the LD setting compared with the SD setting. Detection sensitivities were not different between the SD and the LD settings except between observers 1 and 2 for the detection of micronodules in the thick area. Conclusion LD DT can be effective in nodule detection bigger than 4 mm without a significant decrease in image quality compared with SD DT.
机译:目的确定最佳低剂量(LD)数字断层合成(DT)设置,并将LD DT的图像质量与标准默认(SD)DT的图像质量进行比较。材料和方法通过更改管电压,铜滤波器和剂量比,进行了九种DT设置,以确定LD设置。在DT设置的组合中,确定了提供最低辐射剂量的条件。将80个直径小于1厘米的人造结核(亚厘米结节:40个,小于4毫米微小结节:40个)连接到泡沫聚苯乙烯和体模的隔膜上。其中,有38个结节位于肺部周围(薄区域),有42个结节位于椎旁或sub下区域(厚区域)。四名观察员计数了在厚和薄区域中检测到的结节数量。 SD和LD设置中的检测灵敏度分别计算。对数据进行统计分析。结果最低LD设置是100 kVp,0.3 mm附加铜滤光片和1:5剂量比的组合。 LD和SD设置的有效剂量分别为62μSv和140μSv。与标清设置相比,LD设置的剂量减少了56.7%。 SD和LD设置之间的检测灵敏度没有区别,除了观察员1和2之间用于检测厚区域中的微结节。结论与SD DT相比,LD DT可有效检测大于4 mm的结节,而图像质量不会明显下降。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号