Optimization of chest radiographic imaging parameters: A comparison of image quality and entrance skin dose for digital chest radiography systems

机译:优化胸部放射成像参数:数字胸部放射成像系统的图像质量和入口皮肤剂量的比较

获取原文
获取原文并翻译 | 示例
           

摘要

We studied the performance of three computed radiography and three direct radiography systems with regard to the image noise and entrance skin dose based on a chest phantom. Images were obtained with kVp of 100, 110, and 120 and mA settings of 1, 2, 4, 8, and 10. Significant differences of image noise were found in these digital chest radiography systems (P<0001). Standard deviation was significantly different when the mAs were changed (P<001), but it was independent of the kVp values (P=08-85). Up to 44% of radiation dose could be saved when kVp was reduced from 120 to 100 kVp without compromising image quality.
机译:我们针对基于胸部体模的图像噪声和入射皮肤剂量研究了三种计算机射线照相和三种直接射线照相系统的性能。在kVp为100、110和120且mA设置为1、2、4、8和10的情况下获得图像。在这些数字胸部X线照相系统中发现图像噪声存在显着差异(P <0001)。当mAs改变时,标准偏差显着不同(P <001),但与kVp值无关(P = 08-85)。当kVp从120 kVp降低至100 kVp时,最多可以节省44%的辐射剂量,而不会影响图像质量。