首页> 外文期刊>Nuclear instruments and methods in physics research >Energy Calibration Of Gamma Spectra In Plastic Scintillators Using Compton Kinematics
【24h】

Energy Calibration Of Gamma Spectra In Plastic Scintillators Using Compton Kinematics

机译:利用康普顿运动学对塑料闪烁体中伽马能谱的能量校准

获取原文
获取原文并翻译 | 示例

摘要

This paper describes a simple and practicable method for assigning energy values to gamma-ray pulse-height distributions measured with polyvinyl toluene (PVT) based detectors. It is based upon the characteristic shape of simulated spectra in the region of maximum energy deposition resulting from a single Compton scattering. The validity of this method is demonstrated by applying it to a set of measured NaI(Tl) spectra, and comparing those results to the standard photopeak method of calibrating the same spectra. The method is then applied to a set of measured PVT spectra. It can now be applied to calibration of fielded PVT detectors using two selected gamma-ray sources, which was the initial motivation for this study.
机译:本文介绍了一种简单实用的方法,可将能量值分配给使用基于聚乙烯甲苯(PVT)的探测器测量的伽马射线脉冲高度分布。它基于单个Compton散射产生的最大能量沉积区域中模拟光谱的特征形状。通过将其应用于一组测量的NaI(Tl)光谱,并将这些结果与校准相同光谱的标准光电峰方法进行比较,证明了该方法的有效性。然后将该方法应用于一组测量的PVT光谱。现在可以使用两个选定的伽马射线源将其应用于现场PVT检测器的校准,这是本研究的最初动机。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号