首页> 外文期刊>Instrumentation and Measurement, IEEE Transactions on >A Comparison Between the NIST PJVS-Based Power Standard and the NRC Current- Comparator-Based Power Standard
【24h】

A Comparison Between the NIST PJVS-Based Power Standard and the NRC Current- Comparator-Based Power Standard

机译:基于NIST PJVS的电源标准与基于NRC电流比较器的电源标准之间的比较

获取原文
获取原文并翻译 | 示例

摘要

This paper presents the results of a comparison of active/reactive power meter calibrations between the National Institute of Standards and Technology and National Research Council. The comparison was implemented using a transfer standard consisting of a highly stable commercial sampling-type power/energy meter. Active and reactive power measurements were made at 120 V, 5 A, 50 Hz, and 60 Hz. For active power, the measurements were made at applied current phase angles of 0°, +60°, and -60°. For reactive power, the measurements were made at applied current phase angles of +60°, +90°, -60°, and -90° . The results of the comparison indicate agreement to within the stated uncertainties of the participants.
机译:本文介绍了美国国家标准技术研究院和美国国家研究委员会进行的有功/无功功率计校准的比较结果。使用由高度稳定的商业采样型功率/电能表组成的传输标准来进行比较。在120 V,5 A,50 Hz和60 Hz下进行有功功率和无功功率测量。对于有功功率,在施加的电流相位角为0°,+ 60°和-60°时进行测量。对于无功功率,在施加的电流相角为+ 60°,+ 90°,-60°和-90°时进行测量。比较结果表明在参与者陈述的不确定性范围内是一致的。