摘要

We report the results of a high-energy laser calorimeter comparison conducted by the National Institute of Standards and Technology (NIST), Boulder, Colorado and the U.S. Air Force Primary Standards laboratory (AFPSL), Heath, Ohio. A laser power meter, used as a transfer standard, was calibrated at each laboratory, sequentially, and the measurement results were compared. These measurements were performed at a nominal power of 800 W and a wavelength of 10.6 μm using CO2 lasers. Excellent measurement agreement (1.02 %) was demonstrated, which was well within each of the expanded uncertainties from the two laboratories involved in this comparison.
机译:我们报告了科罗拉多州博尔德的美国国家标准技术研究院(NIST)和俄亥俄州希思市的美国空军主要标准实验室(AFPSL)进行的高能激光量热仪比较的结果。在每个实验室依次校准用作传输标准的激光功率计,并比较测量结果。这些测量是使用CO2激光器在800 W的额定功率和10.6μm的波长下进行的。结果表明,优异的测量一致性(1.02%)完全在此比较所涉及的两个实验室的每个不确定性范围之内。