首页> 外文期刊>Electron Devices, IEEE Transactions on >The Analysis of Potentiometric Flexible Arrayed Urea Biosensor Modified by Graphene Oxide and γ-Fe2O3 Nanoparticles
【24h】

The Analysis of Potentiometric Flexible Arrayed Urea Biosensor Modified by Graphene Oxide and γ-Fe2O3 Nanoparticles

机译:石墨烯和γ-Fe2O3纳米粒子改性电位柔性墨尿素体传感器的分析

获取原文
获取原文并翻译 | 示例

摘要

In this work, we used radio frequency (RF) system and screen printing technology to fabricate urea biosensor based on urease-maghemite nanoparticles ( $gamma $ -Fe2O3 NPs)/graphene oxide (GO)ickel oxide (NiO) sensing membrane on silver conductive wire printed polyethylene terephthalate (PET) substrate. The ${V}$ ${T}$ measurement system was used to investigate the sensing characteristics of urea biosensors based on GO/NiO sensing membrane modified by $gamma $ -Fe2O3 NPs. The sensing membrane was modified by adding different contents of $gamma $ -Fe2O3 NPs. The concentration of $gamma $ -Fe2O3 NPs was 1 mg/mL. When the volume ratio of urease/ $gamma $ -Fe2O3 NPs was 1/1, it had the best sensitivity. The average sensitivity, the linearity, and the limit of detection of (LOD) of the best urea biosensor in this work were 52.36 mV/decade ± 0.22 mV/decade, 0.996, and $7.91~{mu }ext{M}$ , respectively. The effect of temperature on the urea biosensor was investigated from 25 °C to 65 °C. The stability of urea biosensors was evaluated by the effect of hysteresis and long-term storage stability. After 29 days, the average sensitivity remained 80.62% of the initial value. Compared with other urea biosensors, our urea biosensor based on urease- $gamma $ -Fe2O3 NPs/GO/NiO sensing membrane has better average sensitivity.
机译:在这项工作中,我们使用射频(RF)系统和丝网印刷技术来制造基于脲酶-Magemite纳米粒子的尿素生物传感器(<内联公式XMLNS:MML =“http://www.w3.org/1998/math/mathml”xmlns:xlink =“http://www.w3.org/1999/xlink”> $ Gamma $ -fe. 2 O. 3 NPS)/石墨烯氧化物(GO)/氧化镍(NIO)在银导线印刷聚对苯二甲酸乙二醇酯(PET)基材上的传感膜。这<内联公式XMLNS:MML =“http://www.w3.org/1998/math/mathml”xmlns:xlink =“http://www.w3.org/1999/xlink”> $ {v} $ - <内联公式XMLNS:MML =“http://www.w3.org/1998/math/mathml”xmlns:xlink =“http://www.w3.org/1999/xlink”> $ {T} $ 使用测量系统来研究基于GO / NIO传感膜的尿素生物传感器的感测特性<内联公式XMLNS:MML =“http://www.w3.org/1998/math/mathml”xmlns:xlink =“http://www.w3.org/1999/xlink”> $ Gamma $ -fe. 2 O. 3 NPS。通过添加不同的内容物来修饰传感膜<内联公式XMLNS:MML =“http://www.w3.org/1998/math/mathml”xmlns:xlink =“http://www.w3.org/1999/xlink”> $ Gamma $ -fe. 2 O. 3 NPS。浓度<内联公式XMLNS:MML =“http://www.w3.org/1998/math/mathml”xmlns:xlink =“http://www.w3.org/1999/xlink”> $ Gamma $ -fe. 2 O. 3 nps为1mg / ml。当尿素/的体积比/<内联公式XMLNS:MML =“http://www.w3.org/1998/math/mathml”xmlns:xlink =“http://www.w3.org/1999/xlink”> $ Gamma $ -fe. 2 O. 3 NPS是1/1,它具有最好的敏感性。本工作中最佳尿素生物传感器(LOD)的平均灵敏度,线性度和检测极限为52.36 mV /十年±0.22 mV /十年,0.996,和<内联公式XMLNS:MML =“http://www.w3.org/1998/math/mathml”xmlns:xlink =“http://www.w3.org/1999/xlink”> 7.91美元〜{ mu} text {m} $ , 分别。研究了温度对尿素生物传感器的影响,从25℃〜65℃研究。通过滞后和长期储存稳定性的影响评价尿素生物传感器的稳定性。 29天后,平均敏感性占初始值的80.62%。与其他尿素生物传感器相比,我们的尿素生物传感器基于脲酶 - <内联公式XMLNS:MML =“http://www.w3.org/1998/math/mathml”xmlns:xlink =“http://www.w3.org/1999/xlink”> $ Gamma $ -fe. 2 O. 3 NPS / GO / NIO传感膜具有更好的平均灵敏度。

著录项

  • 来源
    《Electron Devices, IEEE Transactions on》 |2020年第11期|5104-5110|共7页
  • 作者单位

    Graduate School of Chemical and Materials Engineering National Yunlin University of Science and Technology Douliou Taiwan;

    Graduate School of Chemical and Materials Engineering National Yunlin University of Science and Technology Douliou Taiwan;

    Tunghai University Taichung Taiwan;

    Graduate School of Electronic Engineering National Yunlin University of Science and Technology Douliou Taiwan;

    Graduate School of Electronic Engineering National Yunlin University of Science and Technology Douliou Taiwan;

    Graduate School of Electronic Engineering National Yunlin University of Science and Technology Douliou Taiwan;

    Graduate School of Chemical and Materials Engineering National Yunlin University of Science and Technology Douliou Taiwan;

    Graduate School of Electronic Engineering National Yunlin University of Science and Technology Douliou Taiwan;

    Graduate School of Electronic Engineering National Yunlin University of Science and Technology Douliou Taiwan;

    Tunghai University Taichung Taiwan;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Iron; Biosensors; Electrodes; Biomembranes; Voltage measurement; Sensitivity;