International Intercomparison of Specific Absorption Rates in a Flat Absorbing Phantom in the Near-Field of Dipole Antennas

机译:偶极子天线近场中扁平吸收体中比吸收率的国际比对

获取原文
获取原文并翻译 | 示例

摘要

This paper reports the results of an international intercomparison of the specific absorption rates (SARs) measured in a flat-bottomed container (flat phantom), filled with human head tissue simulant fluid, placed in the near-field of custom-built dipole antennas operating at 900 and 1800 MHz, respectively. These tests of the reliability of experimental SAR measurements have been conducted as part of a verification of the ways in which wireless phones are tested and certified for compliance with safety standards. The measurements are made using small electric-field probes scanned in the simulant fluid in the phantom to record the spatial SAR distribution. The intercomparison involved a standard flat phantom, antennas, power meters, and RF components being circulated among 15 different governmental and industrial laboratories. At the conclusion of each laboratory's measurements, the following results were communicated to the coordinators: Spatial SAR scans at 900 and 1800 MHz and 1 and 10 g maximum spatial SAR averages for cubic volumes at 900 and 1800 MHz. The overall results, given as meanstandard deviation, are the following: at 900 MHz, 1 g average 7.850.76; 10 g average 5.160.45; at 1800 MHz, 1 g average 18.44 ± 1.65; 10 g average 10.14 ± 0.85, all measured in units of watt per kilogram, per watt of radiated power.
机译:本文报告了在一个装有人体头部组织模拟液的扁平底部容器(扁平体模)中测量的比吸收率(SAR)的国际比对结果,该容器放置在定制偶极子天线的近场中分别为900和1800 MHz这些实验性SAR测量的可靠性测试已作为验证无线电话符合安全标准的测试和验证方法的一部分进行了验证。使用在幻像模拟流体中扫描的小型电场探头进行测量,以记录空间SAR分布。此次比对涉及在15个不同的政府和工业实验室之间流通的标准平面幻象,天线,功率计和RF组件。在每个实验室的测量结束时,将以下结果传达给协调员:900和1800 MHz的空间SAR扫描,以及900和1800 MHz的立方体积的最大空间SAR平均1和10 g。以平均标准偏差给出的总体结果如下:在900 MHz下,1 g平均值为7.850.76。 10克平均5.160.45;在1800 MHz下,1 g平均18.44±1.65; 10克平均值为10.14±0.85,所有均以瓦特/千克,每瓦特的辐射功率为单位进行测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号