首页> 外文期刊>Electromagnetic Compatibility, IEEE Transactions on >EMI Risk Assessment in a Hospital Ward With One and Two Roaming Wireless Transmitters
【24h】

EMI Risk Assessment in a Hospital Ward With One and Two Roaming Wireless Transmitters

机译:带有一个和两个漫游无线发射器的医院病房中的EMI风险评估

获取原文
获取原文并翻译 | 示例

摘要

To control electromagnetic interference, hospitals often specify that mobile transmitters may not be brought any closer than a minimum separation distance (MSD) to an electronic medical device. This paper investigates the risk that the field strength due to mobile transmitters exceeds the immunity level of a medical device. The spatial variation of the field strength can be characterized by the well-known Ricean probability distribution, using the Sabine method to evaluate the parameters. The mobility of transmitters is accounted for by assuming a function for the probability that a transmitter is present at each location throughout the hospital room. The risk of exceeding immunity is estimated with no restriction on the movement of mobile transmitters, and the reduction in risk is estimated when an MSD policy is enforced. Staff may not fully comply with the MSD, so the increase in risk with specified levels of non-compliance is found. It is shown that with some non-compliance the risk of exceeding immunity becomes constant with increasing MSD, and so specifying a larger MSD does not necessarily increase safety.
机译:为了控制电磁干扰,医院通常会规定移动发射机与电子医疗设备之间的距离不得小于最小间隔距离(MSD)。本文研究了由移动发射器引起的场强超过医疗设备的抗扰度水平的风险。场强的空间变化可以通过众所周知的莱斯概率分布来表征,使用Sabine方法评估参数。发射机的可移动性是通过假设某个函数存在于整个病房中每个位置的概率来解决的。估计不受干扰的风险不受移动发射机的移动的限制,并且在实施MSD策略时可以估计降低的风险。员工可能未完全遵守MSD,因此发现特定级别的不遵守行为会增加风险。结果表明,在某些不合规的情况下,随着MSD的增加,超出免疫力的风险将变得恒定,因此指定较大的MSD并不一定会提高安全性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号