Breakage of angioplasty device in coronary artery can cause panic in the catheterization laboratory. These broken fragments may serve as a nidus for thrombus formation; hence, removal of these fragments becomes mandatory. Since the incidence of guidewire fracture during angioplasty are rarely reported, evidence-based approaches are not available for the management of such incidental conditions. Here, we report an interesting case of entrapment and unravelling of guidewire. We successfully retrieved unravelled guidewire using a noncompliant balloon inflated in the guiding catheter. Subsequently, the procedure was completed successfully with an implantation of a stent in the culprit lesion. We are of opinion that this novel technique is quite easy and less cumbersome than other described techniques reported earlier.
机译:冠状动脉血管成形术装置的损坏会在导管实验室引起恐慌。这些破碎的碎片可能充当血栓形成的病菌。因此,必须清除这些碎片。由于很少报道血管成形术中导丝骨折的发生率,因此尚无基于证据的方法可用于此类偶发情况的管理。在这里,我们报告导丝被夹住和散开的有趣情况。我们使用引导导管中膨胀的不顺应性气球成功地取出了散开的导丝。随后,在罪犯病变处植入支架成功地完成了手术。我们认为,与之前报道的其他描述的技术相比,这种新颖的技术非常容易并且不那么麻烦。