Effects of reduced rebreathing time, in spontaneously breathing patients, on respiratory effort and accuracy in cardiac output measurement when using a partial carbon dioxide rebreathing technique: a prospective observational study

机译:使用部分二氧化碳再呼吸技术时,自发呼吸患者减少的再呼吸时间对呼吸努力和心输出量测量准确性的影响:一项前瞻性观察性研究

获取原文
       

摘要

IntroductionNew technology using partial carbon dioxide rebreathing has been developed to measure cardiac output. Because rebreathing increases respiratory effort, we investigated whether a newly developed system with 35 s rebreathing causes a lesser increase in respiratory effort under partial ventilatory support than does the conventional system with 50 s rebreathing. We also investigated whether the shorter rebreathing period affects the accuracy of cardiac output measurement.MethodOnce a total of 13 consecutive post-cardiac-surgery patients had recovered spontaneous breathing under pressure support ventilation, we applied a partial carbon dioxide rebreathing technique with rebreathing of 35 s and 50 s in a random order. We measured minute ventilation, and arterial and mixed venous carbon dioxide tension at the end of the normal breathing period and at the end of the rebreathing periods. We then measured cardiac output using the partial carbon dioxide rebreathing technique with the two rebreathing periods and using thermodilution.ResultsWith both rebreathing systems, minute ventilation increased during rebreathing, as did arterial and mixed venous carbon dioxide tensions. The increases in minute ventilation and arterial carbon dioxide tension were less with 35 s rebreathing than with 50 s rebreathing. The cardiac output measures with both systems correlated acceptably with values obtained with thermodilution.ConclusionWhen patients breathe spontaneously the partial carbon dioxide rebreathing technique increases minute ventilation and arterial carbon dioxide tension, but the effect is less with a shorter rebreathing period. The 35 s rebreathing period yielded cardiac output measurements similar in accuracy to those with 50 s rebreathing.
机译:简介已开发出使用部分二氧化碳呼吸的新技术来测量心输出量。因为重新呼吸增加了呼吸作用,所以我们调查了新开发的35 s呼吸系统在部分通气支持下是否比传统的50 s呼吸系统引起的呼吸力增加量少。我们还研究了较短的呼吸时间是否会影响心输出量测量的准确性。方法一旦总共13例心脏手术后患者在压力支持通气下恢复了自发呼吸,我们采用了部分二氧化碳呼吸技术,并进行了35 s的呼吸并以随机顺序50秒。我们测量了正常呼吸周期结束和再呼吸周期结束时的分钟通气量以及动脉和混合静脉二氧化碳的张力。然后我们使用部分二氧化碳再呼吸技术在两个再呼吸期间并使用热稀释法测量心输出量。结果在两个再呼吸系统中,再呼吸过程中的分钟通气量增加,动脉和混合静脉二氧化碳张力也增加。再呼吸35 s时,分钟通气量和动脉二氧化碳张力的增加要少于再呼吸50 s时。两种系统的心输出量指标与通过热稀释获得的值可以接受地相关。结论当患者自发呼吸时,部分二氧化碳再呼吸技术会增加分钟通气量和动脉二氧化碳张力,但在较短的呼吸时间下效果会较小。 35 s的呼吸周期产生的心输出量测量值与50 s的呼吸具有相似的准确度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号