Reliability of the Partial CO_2 Rebreathing Technique for Measurement of Cardiac Output

机译:部分CO_2再呼吸技术在心输出量测量中的可靠性

获取原文
获取原文并翻译 | 示例

摘要

The partial CO_2 rebreathing technique has been suggested to accurately estimate cardiac output (Q_t) even in the presence of lung injury. Using a computer-based gas exchange mesurement system and graphical and textual programming languages, we evaluated the reliability of this noninvasive technique for measurement of cardiac output in the setting of lung atelectasis and lung injury. The error of estimating Q_t with this method correlates with the right to leftrnintrapulmonary shunt fraction (Q_s / Q_t). We conclude that this technique is not reliable in the setting of lung injury, but is useful for the mechanically ventilated patient with normalrnQ_s/Q_t.
机译:已建议使用部分CO_2再呼吸技术,即使在存在肺损伤的情况下也可以准确估算心输出量(Q_t)。使用基于计算机的气体交换测量系统以及图形和文本编程语言,我们评估了这种非侵入性技术在肺不张和肺损伤情况下测量心输出量的可靠性。用这种方法估算Q_t的误差与右肺内分流分数从左到右(Q_s / Q_t)相关。我们得出的结论是,该技术在肺部损伤的治疗中并不可靠,但对rnQ_s / Q_t正常的机械通气患者很有用。