Validation of dosimetric field matching accuracy from proton therapy using a robotic patient positioning system

机译:使用机器人患者定位系统验证质子治疗的剂量场匹配精度

获取原文
           

摘要

Large area, shallow fields are well suited to proton therapy. However, due to beam production limitations, such volumes typically require multiple matched fields. This is problematic due to the relatively narrow beam penumbra at shallow depths compared to electron and photon beams. Therefore, highly accurate dose planning and delivery is required. As the dose delivery includes shifting the patient for matched fields, accuracy at the 1–2 millimeter level in patient positioning is also required. This study investigates the dosimetric accuracy of such proton field matching by an innovative robotic patient positioner system (RPPS). The dosimetric comparisons were made between treatment planning system calculations, radiographic film and ionization chamber measurements. The results indicated good agreement amongst the methods and suggest that proton field matching by a RPPS is accurate and efficient. PACS number: 87.55.km
机译:大面积的浅场非常适合质子治疗。但是,由于束产生的限制,这种体积通常需要多个匹配的场。这是有问题的,因为与电子束和光子束相比,在浅深度处的束半影较窄。因此,需要高度精确的剂量计划和递送。由于剂量输送包括将患者移至相匹配的视野,因此在患者定位中还要求精度为1-2毫米。这项研究调查了通过创新的机器人病人定位器系统(RPPS)进行此类质子场匹配的剂量学准确性。在治疗计划系统计算,射线照相胶片和电离室测量之间进行剂量学比较。结果表明这些方法之间的良好一致性,并表明通过RPPS进行质子场匹配是准确和高效的。 PACS号码:87.55.km