Imaging cardiac sympathetic innervation with MIBG: linear conversion of the heart-to-mediastinum ratio between different collimators

机译:用MIBG成像心脏交感神经:不同准直仪之间心对纵隔比率的线性转换

获取原文
           

摘要

Abstract BackgroundThe heart-to-mediastinum (H/M) ratio is a commonly used parameter to measure cardiac I-123 metaiodobenzylguanidine (MIBG) uptake. Since the H/M ratio is substantially influenced by the collimator type, we investigated whether an empirical linear conversion of H/M ratios between camera systems with low-energy (LE) and medium-energy (ME) collimator is possible.MethodsWe included 18 patients with parkinsonism who were referred to one of the two participating molecular imaging facilities for the evaluation of cardiac sympathetic innervation by MIBG scintigraphy. Two consecutive planar image datasets were acquired with LE and ME collimators at 4 h after MIBG administration. Linear regression analyses were performed to describe the association between the H/M ratios gained with both collimator settings, and the accuracy of a linear transfer of the H/M ratio between collimators and across centers was assessed using a leave-one-out procedure.ResultsH/M ratios acquired with LE and ME collimators showed a strong linear relationship both within each imaging facility ( R 2 = 0.99, p 0.001 and R 2 = 0.90, p 0.001) and across centers (H/M-LE = 0.41 × H/M-ME + 0.63, R 2 = 0.97, p 0.001). A linear conversion of H/M ratios between collimators and across centers was estimated to be very accurate (mean absolute error 0.05 ± 0.04; mean relative absolute error 3.2 ± 2.6%).ConclusionsThe present study demonstrates that a simple linear conversion of H/M ratios acquired with different collimators is possible with high accuracy. This should greatly facilitate the exchange of normative data between settings and pooling of data from different institutions.
机译:摘要背景心/纵隔(H / M)比是测量心脏I-123异碘苄基胍(MIBG)摄取的常用参数。由于H / M比率受准直器类型的影响很大,因此我们研究了低能(LE)和中能(ME)准直器的摄像系统之间H / M比率的经验线性转换是否可行。方法我们包括18个帕金森病患者被转介到两个参与的分子影像学设施之一,以通过MIBG闪烁显像技术评估心脏交感神经。 MIBG给药后4小时,使用LE和ME准直仪采集了两个连续的平面图像数据集。进行线性回归分析以描述在两种准直仪设置下获得的H / M比率之间的关联,并使用留一法程序评估了准直仪之间和跨中心的H / M比率线性传递的准确性。结果用LE和ME准直仪获得的H / M比在每个成像设备内(R 2 = 0.99,p <0.001和R 2 = 0.90,p <0.001)以及整个中心(H / M-LE = 0.41)都显示出很强的线性关系×H / M-ME + 0.63,R 2 = 0.97,p <0.001)。估计准直仪之间和跨中心的H / M比率的线性转换非常准确(平均绝对误差0.05±0.04;平均相对绝对误差3.2±2.6%)。结论本研究表明,H / M的简单线性转换使用不同的准直器获得的光圈比可以高精度地获得。这将极大地促进设置之间的规范性数据的交换和来自不同机构的数据池。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号