An improved method for estimating the heart-to-mediastinum ratio from cardiac sympathetic nerve imaging with low-energy high-resolution collimators

机译:一种用低能量高分辨率准直器从心脏交感神经成像估计心-纵隔比率的改进方法

获取原文
获取原文并翻译 | 示例
           

摘要

Background. Septal penetration causes underestimation of the heart-to-mediastinum (H/M) ratio in cardiac 123I- metaiodobenzylguanidine (MIBG) imaging with a low-energy high-resolution (LEHR) collimator. We aimed to improve the method of estimating the H/M ratios using the LEHR collimator. Methods and Results. 4 hours after 123I-MIBG injection, 40 patients were imaged successively with the medium-energy (ME) and LEHR collimators using gamma cameras having 3/8-inch crystals. Severe underestimation of the H/M ratios was observed with the LEHR collimator when compared to the ME collimator. Narrowing the energy window width did not reduce the underestimation. Application of 123I-dual-window (IDW) correction using a narrow or wide subwindow reduced the underestimation substantially but not entirely. The H/M ratios estimated from the LEHR images with or without IDW correction were corrected based on their correlations with the ratios estimated from the ME images. This empiric correction removed systematic underestimation, and residual errors were reduced when the H/M ratios after IDW correction were converted using the empiric equation. The conversion equation was successfully applied to the correction of the H/M ratios determined in another 40 patients using a 5/8-inch crystal. Conclusions. In estimating the H/M ratios using an LEHR collimator, empiric correction combined with IDW correction improves concordance with ME-based values in comparison with empiric correction alone.
机译:背景。间隔穿透会导致低能量高分辨率(LEHR)准直仪在心脏123I-碘碘苄基胍(MIBG)成像中低估心脏与纵隔(H / M)的比率。我们旨在改进使用LEHR准直仪估算H / M比率的方法。方法和结果。注射123I-MIBG后4小时,使用具有3/8英寸晶体的伽马相机,用中能(ME)和LEHR准直仪对40位患者进行连续成像。与ME准直仪相比,使用LEHR准直仪观察到H / M比的严重低估。缩小能量窗口宽度并不能减少低估。使用较窄或较宽子窗口的123I双窗口(IDW)校正的应用可大大降低低估,但不能完全消除。根据LEHR图像与IDW校正之间或不与IDW校正之间的相关性,可以校正从LEHR图像估计的H / M比值。该经验校正消除了系统的低估,当使用经验方程转换IDW校正后的H / M比时,可以减少残留误差。该转换方程已成功地应用于校正另一位40例使用5/8英寸晶体的H / M比值。结论。在使用LEHR准直器估算H / M比率时,与单独的经验校正相比,经验校正与IDW校正相结合可改善与基于ME的值的一致性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号