We implement a test bed allowing the characterisation of cantilevers used in non-contact atomic force microscopy under controlled atmospheric conditions prior to using the cantilevers in a force microscope.The effective Q-factor of two types of cantilevers is measured as a function of the ambient pressure varied from 10~(-8) mbar to normal pressure.The Q-factor is found to be almost constant up to a pressure in the range of 10~(-2) to 10~(-1) mbar and then decreases by about three ordes of magnitude when increasing the pressure further to normal pressure.The pressure dependence of the effective Q-factor is approximated by analytical models based on the intrinsic damping of the cantilever, mounting losses and pressure dependent terms accounting for damping in the molecular flow and viscous regimes.The experimental data can be well described by the models in both regimes.
机译:我们安装了一个测试台,可以在受控显微镜下表征非接触式原子力显微镜中使用的悬臂,然后再在力显微镜中使用悬臂。两种悬臂的有效Q因子随环境的变化进行测量压力在10〜(-8)mbar至常压之间变化.Q值在10〜(-2)到10〜(-1)mbar的压力范围内几乎恒定,然后降低当压力进一步增加到常压时大约为三个数量级。有效Q因子的压力依赖性通过基于悬臂固有阻尼,安装损耗和考虑分子流阻尼的压力相关项的解析模型来近似实验数据可以用两种模式的模型很好地描述。