首页> 外文期刊>IEEE Transactions on Applied Superconductivity >Repeatability of critical-current measurements on Nb/sub 3/Sn and Nb-Ti wires
【24h】

Repeatability of critical-current measurements on Nb/sub 3/Sn and Nb-Ti wires

机译:Nb / sub 3 / Sn和Nb-Ti导线上临界电流测量的可重复性

获取原文
获取原文并翻译 | 示例

摘要

A varying degree of repeatability has been observed in critical-current (I/sub c/) measurements of Nb/sub 3/Sn and Nb-Ti wires as a function of the number of thermal cycles from room temperature to 4 K. The increase of I/sub c/ between the first and second thermal cycle can be 1% to 2% at 12 T for Nb/sub 3/Sn wires. This was observed on Nb/sub 3/Sn wire by all four laboratories that participated in a recent interlaboratory comparison conducted in the International Thermonuclear Experimental Reactor (ITER) project. These data indicate that if I/sub c/ changes beyond the error limits, it increases fairly monotonically with thermal cycling until it eventually saturates. In contrast, the I/sub c/ of a Nb-Ti wire is very repeatable with thermal cycling. This suggests that the effect on the Nb/sub 3/Sn wire is due to its strain sensitivity. Most of these data were taken with the sample on a Ti-6Al-4V measurement mandrel. This study also investigated the repeatability of I/sub c/ measurements using other mandrel materials. The increase in I/sub c/ of Nb/sub 3/Sn wire could enhance the performance of some applications. However, the lack of repeatability in I/sub c/ measurements on Nb/sub 3/Sn wires is a limitation in precise interlaboratory comparisons.
机译:在Nb / sub 3 / Sn和Nb-Ti导线的临界电流(I / sub c /)测量中,观察到了不同程度的重复性,这是从室温到4 K的热循环次数的函数。对于Nb / sub 3 / Sn线,在第一个热循环和第二个热循环之间的I / sub c /的12 T时的I / sub c /的百分比可以为1%到2%。参与最近在国际热核实验反应堆(ITER)项目中进行的实验室间比较的所有四个实验室均在Nb / sub 3 / Sn线上观察到了这一点。这些数据表明,如果I / sub c /的变化超出误差限制,则它会随着热循环而相当单调地增加,直到最终饱和。相比之下,Nb-Ti线的I / sub c /在热循环中非常可重复。这表明对Nb / sub 3 / Sn焊丝的影响归因于其应变敏感性。这些数据大多数是在Ti-6Al-4V测量心轴上与样品一起采集的。这项研究还调查了使用其他心轴材料进行I / sub c /测量的可重复性。 Nb / sub 3 / Sn线的I / sub c /的增加可以增强某些应用程序的性能。但是,在Nb / sub 3 / Sn导线上进行I / sub c /测量时,缺乏可重复性,这是实验室间精确比较的一个局限。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号