We describe a low field, bench-top MRI system for small samples, based on a permanent magnet. Signals are received at 830 kHz using a tuned SQUID magnetometer, cooled in a modified liquid helium cryostat. The SQUID input circuit has an intrinsic $Q$-factor of 28,000, so this is damped by flux-locking electronics and an additional feedback loop to give an effective $Q$ of between 40 and 200 for imaging. The resonant frequency is adjusted by a control rod coupled to a trimmer capacitor mounted in the liquid helium volume. Images of a test object acquired using the cooled receiver exhibit up to two-fold SNR gains in regions close to the sensor, compared to an equivalent room temperature coil, with the noise level dominated by losses coupled from the magnet pole faces.
机译:我们描述了一种基于永磁体的小样品低场台式MRI系统。使用调谐的SQUID磁力计以830 kHz接收信号,并在改良的液氦低温恒温器中冷却。 SQUID输入电路具有28,000的固有$ Q $因数,因此可通过磁通量锁定电子设备和附加的反馈环路来衰减,以使成像的有效QQ $在40到200之间。谐振频率由与安装在液氦体积中的微调电容器相连的控制杆调节。与等效的室温线圈相比,使用冷却的接收器采集的测试对象的图像在靠近传感器的区域显示出高达两倍的SNR增益,而噪声水平主要由磁极面耦合的损耗决定。