Analytical quality assessment and method comparison of immunoassaysfor the measurement of serum cobalamin and folate in dogs andcats

机译:免疫测定的分析质量评估和方法比较用于狗和狗血清中钴胺素和叶酸的测定猫

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Serum cobalamin and folate are often measured in cats and dogs as part oflaboratory testing for intestinal disease, small intestinal dysbiosis, orexocrine pancreatic deficiency. We performed an analytical validation of humanimmunoassays for cobalamin and folate measurement (AIA-900 analyzer, TosohBioscience) and compared results with those obtained using chemiluminescenceassays (Immulite 2000 analyzer, Siemens Medical Solutions Diagnostics).Accuracy, precision, total observable error (TE %), and σ valueswere calculated for the immunoassays. Correlation and agreement were evaluatedwith Deming regression, Passing–Bablok regression, and Bland–Altman analysis.Cobalamin intra-assay and inter-assay CVs were 1.8–9.3% and 2.6–6.8%,respectively. Folate intra-assay and inter-assay CVs were 1.5–9.1% and 3.4–8.1%,respectively. TE (%) were ≤19 and ≤31 for cobalamin and folate,respectively. Sigma values were 3.60–11.50 for cobalamin and 2.90–7.50 forfolate. Regression analysis demonstrated very high or high correlations forcobalamin [ = 0.98 (dogs), 0.97 (cats)] and folate[ = 0.88 (dogs), 0.92 (cats)] but Bland–Altman analysisrevealed poor agreement for both. The immunoassays had good analyticalperformance for measuring cobalamin and folate in both species. Results obtainedby the 2 analyzers cannot be used interchangeably and should be interpretedusing instrument-specific reference intervals. Further studies are required toestablish immunoassay-specific reference intervals and to evaluate thediagnostic performance and clinical utility of the analyzer for theseanalytes.
机译:血清钴胺素和叶酸通常在猫和狗中测量,作为一部分实验室检查是否有肠道疾病,小肠营养不良或外分泌型胰腺缺乏症。我们对人进行了分析验证钴胺素和叶酸测定的免疫测定法(AIA-900分析仪,Tosoh生物科学),并将结果与​​化学发光法比较分析(Immulite 2000分析仪,西门子医疗解决方案诊断公司)。准确性,精度,总可观察误差(TE%)和σ值进行免疫测定。评估相关性和一致性包括Deming回归,Passing–Bablok回归和Bland–Altman分析。Cobalamin批内和批间CV分别为1.8–9.3%和2.6–6.8%,分别。叶内测定内和测定间CV分别为1.5–9.1%和3.4–8.1%,分别。钴胺素和叶酸的TE(%)≤19和≤31,分别。钴胺素的西格玛值是3.60–11.50,而钴胺素的西格玛值是2.90–7.50叶酸。回归分析表明与钴胺素[= 0.98(狗),0.97(猫)]和叶酸[= 0.88(狗),0.92(猫)],但Bland–Altman分析透露了双方的不佳协议。免疫分析具有良好的分析能力在两个物种中测量钴胺素和叶酸的性能。获得的结果由2个分析器组成的分析仪不能互换使用,应解释为使用特定于仪器的参考间隔。需要进一步研究建立免疫测定特异性参考区间并评估分析仪的诊断性能和临床实用性分析物。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号