首页> 外文期刊>Journal of Veterinary Diagnostic Investigation >Analytical quality assessment and method comparison of immunoassays for the measurement of serum cobalamin and folate in dogs and cats
【24h】

Analytical quality assessment and method comparison of immunoassays for the measurement of serum cobalamin and folate in dogs and cats

机译:免疫测定血清钴胺和叶猫叶酸血清叶酸和叶酸叶酸血清的分析质量评估及方法比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Serum cobalamin and folate are often measured in cats and dogs as part of laboratory testing for intestinal disease, small intestinal dysbiosis, or exocrine pancreatic deficiency. We performed an analytical validation of human immunoassays for cobalamin and folate measurement (AIA-900 analyzer, Tosoh Bioscience) and compared results with those obtained using chemiluminescence assays (Immulite 2000 analyzer, Siemens Medical Solutions Diagnostics). Accuracy, precision, total observable error (TE_(obs)%), and σ values were calculated for the immunoassays. Correlation and agreement were evaluated with Deming regression, Passing–Bablok regression, and Bland–Altman analysis. Cobalamin intra-assay and inter-assay CVs were 1.8–9.3% and 2.6–6.8%, respectively. Folate intra-assay and inter-assay CVs were 1.5–9.1% and 3.4–8.1%, respectively. TE_(obs)(%) were ≤19 and ≤31 for cobalamin and folate, respectively. Sigma values were 3.60–11.50 for cobalamin and 2.90–7.50 for folate. Regression analysis demonstrated very high or high correlations for cobalamin [ r = 0.98 (dogs), 0.97 (cats)] and folate [ r = 0.88 (dogs), 0.92 (cats)] but Bland–Altman analysis revealed poor agreement for both. The immunoassays had good analytical performance for measuring cobalamin and folate in both species. Results obtained by the 2 analyzers cannot be used interchangeably and should be interpreted using instrument-specific reference intervals. Further studies are required to establish immunoassay-specific reference intervals and to evaluate the diagnostic performance and clinical utility of the analyzer for these analytes.
机译:血清钴胺和叶酸通常以猫和狗衡量,作为肠疾病,小肠消带病或外胚层胰腺缺乏的实验室检测的一部分。我们对钴胺素和叶酸测量(AIA-900分析仪,Tosoh Bioscience)进行了人类免疫测定的分析验证,并将结果与​​使用化学发光测定(Immulite 2000分析仪,西门子医疗解决方案诊断)进行了比较的结果。为免疫测定计算精度,精度,总观察到误差(TE_(OB)%)和σ值。使用脱亮回归,通过 - Bablok回归和Bland-Altman分析评估了相关性和协议。 Cobalamin分析和测定间CV分别为1.8-9.3%和2.6-6.8%。叶酸分析和测定间CV分别为1.5-9.1%和3.4-8.1%。 TE_(OB)(%)分别为钴胺和叶酸的≤19和≤31。 Sigma值为Cobalamin的3.60-11.50,叶酸的2.90-7.50。回归分析表现出对钴胺的非常高或高的相关性[r = 0.98(狗),0.97(猫)和叶酸[r = 0.88(狗),0.92(猫)],但Bland-Altman分析揭示了两者的差。免疫测定具有良好的分析性能,用于测量两种物种中的钴胺素和叶酸。由2分析仪获得的结果不能互换使用,并且应该使用特定仪器的参考间隔来解释。需要进一步的研究来建立免疫测定特异性的参考间隔,并评估这些分析物的分析仪的诊断性能和临床效用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号