首页> 美国卫生研究院文献>Biomedical Optics Express >Optimization of fluorescent imaging in the operating room through pulsed acquisition and gating to ambient background cycling
【2h】

Optimization of fluorescent imaging in the operating room through pulsed acquisition and gating to ambient background cycling

机译:通过脉冲采集和门控到环境背景循环来优化手术室中的荧光成像

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The design of fluorescence imaging instruments for surgical guidance is rapidly evolving, and a key issue is to efficiently capture signals with high ambient room lighting. Here, we introduce a novel time-gated approach to fluorescence imaging synchronizing acquisition to the 120 Hz light of the room, with pulsed LED excitation and gated ICCD detection. It is shown that under bright ambient room light this technique allows for the detection of physiologically relevant nanomolar fluorophore concentrations, and in particular reduces the light fluctuations present from the room lights, making low concentration measurements more reliable. This is particularly relevant for the light bands near 700nm that are more dominated by ambient lights.
机译:用于外科手术指导的荧光成像仪器的设计正在迅速发展,而关键问题是在高室内照明条件下有效捕获信号。在这里,我们介绍了一种新颖的时间门控方法,通过脉冲LED激发和门控ICCD检测,将荧光同步采集到房间的120 Hz光。结果表明,在明亮的室内光线下,该技术可以检测生理上相关的纳摩尔荧光团浓度,特别是减少室内光线引起的光波动,从而使低浓度测量更加可靠。这对于700nm附近的波段尤其重要,该波段更多地被环境光所控制。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号