Reproducibility of the Heidelberg retinal flowmeter in determining low perfusion areas in peripapillary retina

机译:海德堡视网膜流量计在确定乳头周围视网膜低灌注区域时的可重复性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Aim: To evaluate the interobserver variability and retest reproducibility of confocal scanning laser Doppler flowmeter in measuring capillary perfusion of the peripapillary retina.>Methods: Blood flow measurements were performed in one eye of 10 normal subjects by two investigators on two different days (visits). Five separate measurements of the peripapillary blood flow parameters were recorded by each observer at each visit. The Heidelberg retina flowmeter was used to record capillary perfusion in a 2560×640 μm area of the superotemporal peripapillary region and pixel by pixel analysis was done from an area adjacent to the optic disc, with a minimum of 1600 pixels. The percentage of pixels with less than 1 arbitrary unit of flow (no flow) and 10, 25, 50, 75, 90th percentiles of flow values was calculated. Interobserver measurements were compared by paired t test. Intraclass correlations (ICC) were used to determine the interobserver variability and retest reproducibility of the measurements. Intrasession coefficients of variations (CV) were also calculated.>Results: There were no statistically significant differences between the two observers for all measurements and between visits for the percentage of pixels with no flow. ICC was 66% (range 57.09%–77.1%) for pixels with no flow. For the 10, 25, 50, 75, 90th percentiles of flow the ICC was 63.07% (53.91%–77.81%), 71.3% (64.23%–80.85%), 72.61% (66.02%–78.96%), 65.86% (58.53%–74.77%), and 60.05% (54.34%–70.06%), respectively. CV was 16.59%, 11.47%, 9.32%, 9.03%, 11.58%, and 16.05% for the percentage of no flow pixels and the 10, 25, 50, 75, 90th percentiles of flow, respectively.>Conclusions: The Heidelberg retinal flowmeter allows reproducible measurements of all levels of capillary perfusion areas when pixel by pixel analysis is used.
机译:>目的:为了评估共聚焦扫描激光多普勒流量计在测量乳头周围视网膜毛细血管灌注中的观察者间差异和重测重现性。>方法:由两名研究人员在两个不同的日期(访问)进行10个正常受试者的研究。每个观察者在每次访视时记录五次单独的乳头周血流参数测量值。使用海德堡视网膜流量计记录颞上周乳头状区域2560×640μm区域的毛细血管灌注,并从与视盘相邻的区域(至少1600个像素)逐个像素进行像素分析。计算了具有小于1个任意流量单位(无流量)和流量值的第10、25、50、75、90个百分位的像素百分比。观察者间的测量通过配对t检验进行比较。类内相关性(ICC)用于确定观察者间的变异性和测量的重测重现性。 >结果:两个观察者之间的所有测量结果和访问之间的无流量像素百分比在统计学上均无显着差异。无流像素的ICC为66%(范围57.09%–77.1%)。对于10、25、50、75、90%的流量,ICC分别为63.07%(53.91%–77.81%),71.3%(64.23%–80.85%),72.61%(66.02%–78.96%),65.86%( 58.53%–74.77%)和60.05%(54.34%–70.06%)。无流量像素的百分比和流量的第10、25、50、75、90个百分位数的CV分别为16.59%,11.47%,9.32%,9.03%,11.58%和16.05%。>结论:< / strong>使用逐像素分析时,海德堡视网膜流量计可对所有水平的毛细血管灌注区域进行可重复的测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号