Reproducibility of the Heidelberg retinal flowmeter in determining low perfusion areas in peripapillary retina.

机译:海德堡视网膜流量计在确定乳头周围视网膜低灌注区域时的可重复性。

获取原文
获取原文并翻译 | 示例
       

摘要

AIM: To evaluate the interobserver variability and retest reproducibility of confocal scanning laser Doppler flowmeter in measuring capillary perfusion of the peripapillary retina. METHODS: Blood flow measurements were performed in one eye of 10 normal subjects by two investigators on two different days (visits). Five separate measurements of the peripapillary blood flow parameters were recorded by each observer at each visit. The Heidelberg retina flowmeter was used to record capillary perfusion in a 2560x640 microm area of the superotemporal peripapillary region and pixel by pixel analysis was done from an area adjacent to the optic disc, with a minimum of 1600 pixels. The percentage of pixels with less than 1 arbitrary unit of flow (no flow) and 10, 25, 50, 75, 90th percentiles of flow values was calculated. Interobserver measurements were compared by paired t test. Intraclass correlations (ICC) were used to determine the interobserver variability and retest reproducibility of the measurements. Intrasession coefficients of variations (CV) were also calculated. RESULTS: There were no statistically significant differences between the two observers for all measurements and between visits for the percentage of pixels with no flow. ICC was 66% (range 57.09%-77.1%) for pixels with no flow. For the 10, 25, 50, 75, 90th percentiles of flow the ICC was 63.07% (53.91%-77.81%), 71.3% (64.23%-80.85%), 72.61% (66.02%-78.96%), 65.86% (58.53%-74.77%), and 60.05% (54.34%-70.06%), respectively. CV was 16.59%, 11.47%, 9.32%, 9.03%, 11.58%, and 16.05% for the percentage of no flow pixels and the 10, 25, 50, 75, 90th percentiles of flow, respectively. CONCLUSIONS: The Heidelberg retinal flowmeter allows reproducible measurements of all levels of capillary perfusion areas when pixel by pixel analysis is used.
机译:目的:评估共焦扫描激光多普勒流量计在测量周围乳头视网膜毛细血管灌注中的观察者间差异和重测重现性。方法:由两名研究人员在两天不同的时间(拜访)对10名正常受试者的一只眼睛进行血流量测量。每个观察者在每次访问时记录五次单独的乳头周血流参数测量值。使用Heidelberg视网膜流量计记录颞上周乳头状区域2560x640微米区域的毛细血管灌注,并从与视盘相邻的区域(至少1600个像素)逐像素进行像素分析。计算了具有小于1个任意流量单位(无流量)和10、25、50、75、90个百分位流量值的像素百分比。观察者间的测量通过配对t检验进行比较。类内相关性(ICC)用于确定观察者间的变异性和测量值的重测重现性。还计算了会间变异系数(CV)。结果:在两个观察者之间,所有测量值和两次访问之间的无流量像素百分比在统计学上均无显着差异。对于没有流量的像素,ICC为66%(范围57.09%-77.1%)。对于第10、25、50、75、90%的流量,ICC为63.07%(53.91%-77.81%),71.3%(64.23%-80.85%),72.61%(66.02%-78.96%),65.86%( 58.53%-74.77%)和60.05%(54.34%-70.06%)。对于无流量像素的百分比以及流量的第10、25、50、75、90个百分点,CV分别为16.59%,11.47%,9.32%,9.03%,11.58%和16.05%。结论:使用逐像素分析时,海德堡视网膜流量计可对所有水平的毛细血管灌注区域进行可重复的测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号