Evaluation of the arterial blood pressure of dogs by two noninvasive methods.

机译:通过两种非侵入性方法评估狗的动脉血压。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

The determination of the arterial blood pressure was done on 12 healthy mixed breed dogs in both the anesthetized and the conscious state, to evaluate two instruments (Doppler flow detector and infrasonde D4000), in their ability to indirectly determine arterial blood pressure. The coefficients of variation were higher with indirect methods when compared with the results obtained by cannulation. These coefficients were lower with the Doppler flow detector. The correlation study showed that both apparatuses were reliable in most situations. The infrasonde D4000 was more accurate than the Doppler in the conscious animals. However the results showed a lack of precision in hypertensive conscious dogs. The diastolic arterial blood pressure was particularly precise in the case of the anesthetized hypotensive dogs. Its sensitivity allowed it to register muscle movement artifacts. The Doppler flow detector showed less variation and was particularly accurate in both anesthetized and conscious hypertensive dogs. Its sensitivity allowed artifact movement sounds to be detected. The Doppler should be used in quiet surroundings or earphones should be worn by the evaluator. Some form of restraint is needed with the use of both instruments. Even if the correlations with the direct arterial blood pressure values were better with the infrasonde D4000, greater variations were found in the individual readings. The Doppler instrument represents in the hands of the investigators a better instrument for routine monitoring of blood pressure in the dog.
机译:在麻醉和清醒状态下,对12只健康的混血狗进行动脉血压的测定,以评估两种仪器(多普勒血流检测仪和Insonsonde D4000)间接测定动脉血压的能力。与通过插管获得的结果相比,间接方法的变异系数更高。这些系数在多普勒血流检测器中较低。相关性研究表明,两种设备在大多数情况下都是可靠的。在有意识的动物中,下探D4000比多普勒更精确。但是,结果表明,高血压意识清醒的犬缺乏精确性。在麻醉的低血压犬的情况下,舒张动脉血压特别精确。它的灵敏度使其可以记录肌肉运动伪影。多普勒血流检测器显示出较小的变化,并且在麻醉和清醒的高血压犬中特别准确。它的灵敏度使得可以检测到伪影移动声音。多普勒仪应在安静的环境中使用,或评估人员应佩戴耳机。使用这两种工具都需要某种形式的约束。即使使用超声波探空仪D4000与直接动脉血压值的相关性更好,在各个读数中也会发现更大的差异。多普勒仪器在研究人员的手中代表了一种更好的仪器,用于常规监测狗的血压。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号