首页> 外文期刊>Hypertension: An Official Journal of the American Heart Association >Evaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure.
【24h】

Evaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure.

机译:对无创血压监测设备Spacelabs 90202和90207与静止和非卧床24小时动脉内血压的评估。

获取原文
   

获取外文期刊封面封底 >>

       

摘要

This study evaluated the accuracy of blood pressure values provided by the Spacelabs 90202 and 90207 devices in comparison with intra-arterial recording in 19 subjects at rest and in nine subjects in ambulatory conditions (Oxford method). At rest Spacelabs monitors reflected intra-arterial systolic blood pressure values very closely but overestimated to a considerable extent intra-arterial diastolic blood pressure (Spacelabs-intra-arterial differences, -0.8 +/- 9.2, NS, and 9.1 +/- 8.8 mm Hg, p less than 0.01, for systolic and diastolic blood pressures, respectively). In ambulatory conditions Spacelabs-intra-arterial average differences in 24-hour values were +0.4 +/- 5.1 mm Hg for systolic blood pressure (NS) and +14.0 +/- 2.9 mm Hg for diastolic blood pressure (p less than 0.01) when group data were considered. The performance of both Spacelabs devices was worse when assessed in individual subjects or for each hourly interval. In spite of these differences between noninvasive and intra-arterial absolute blood pressure values, however, Spacelabs 90202 and 90207 monitors were able to faithfully reflect directional hour-to-hour changes in intra-arterial blood pressure (chi 2 = 18.2 and chi 2 = 23.1 for systolic and diastolic blood pressures, respectively, p less than 0.01). No differences were found between the performance of the two Spacelabs devices. Thus, although the absolute accuracy of blood pressure values provided by these monitors in ambulatory subjects is still limited, they seem to be suitable for studies aimed at assessing 24-hour blood pressure profiles quantitatively as well as qualitatively.
机译:这项研究评估了Spacelabs 90202和90207设备提供的血压值与19位静止受试者和9位非卧床受试者的动脉内记录相比的准确性(牛津法)。在静止状态下,Spacelabs监护仪非常紧密地反映了动脉内的收缩压值,但在相当大的程度上高估了动脉内的舒张压(Spacelabs的动脉内差异为-0.8 +/- 9.2,NS和9.1 +/- 8.8 mm Hg,p分别小于收缩压和舒张压的0.01。在非卧床状态下,Spacelabs的动脉内收缩压(NS)在24小时内的平均差异为+0.4 +/- 5.1 mm Hg,舒张压为+14.0 +/- 2.9 mm Hg(p小于0.01)当考虑组数据时。在个别受试者中或每个小时间隔进行评估时,两个Spacelabs设备的性能均较差。尽管在无创和动脉内绝对血压值之间存在这些差异,但是,Spacelabs 90202和90207监护仪仍能够忠实地反映出动脉内血压每小时的方向变化(chi 2 = 18.2和chi 2 =对于收缩压和舒张压分别为23.1,p小于0.01)。在两个Spacelabs设备的性能之间未发现差异。因此,尽管这些监视器在非卧床受试者中提供的血压值的绝对准确性仍然受到限制,但它们似乎适合于旨在定量和定性评估24小时血压曲线的研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号