首页> 中文期刊> 《环境监测管理与技术》 >离子色谱法测定水中溴离子与碘离子

离子色谱法测定水中溴离子与碘离子

         

摘要

采用离子色谱法测定水中溴离子与碘离子,选择AS19阴离子交换分离柱、KOH淋洗液自动发生器、抑制型电导检测.溴离子与碘离子分别在14.8μg/L ~ 100 mg/L与30.4 μg/L~ 100 mg/L范围内线性良好,检出限分别为3.7 μg/L与7.6μg/L,环境水样平行测定的RSD分别为0.2%~1.5%与0.3% ~1.6%,两个质量浓度水平的加标回收率分别为91.8% ~105%与89.5%~93.4%.%An ion chromatography method for determining Br~ and I" in water was studied. The separation of the anions was achieved on an IonPac AS19 Column with KOH as eluent,and the detection was performed by a conductivity detection mode. Good linearity were obtained within the concentration range of 14. 8 μg/L~ 100 mg/L for Br- and 30.4 μg/L~100 mg/L for I-. The method detection limits of Br- and I- were 3.7 μg/L and 7.6 μg/L, respectively. The RSD of the samples were 0.2%~1.5% (Br-) and 0. 3%~1.6% (I-).The recoveries of Br- in water samples were 91. 8%~105% ,and the recoveries of I- were 89.5%~93.4%.

著录项

相似文献

  • 中文文献
  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号