ASTM F2624-Evaluating Alternate Fixtures for Flexion/Extention, Lateral Bending, and Axial Rotation Testing of Extra Discal Motion Preserving Implants

机译:ASTM F2624-评估用于保留额外椎间盘运动的植入物的屈曲/伸展,横向弯曲和轴向旋转测试的替代固定装置

获取原文
获取原文并翻译 | 示例

摘要

ASTM F2624-07 prescribes applying flexion/extension (FE), lateral bending (LB), and axial rotation (AR) moments to test and evaluate extra discal motion preserving implants (EDMPI), but parameters such as fixture design and load parameters are left to the user to define. A set of fixtures to conduct FE, LB, and AR testing rotation of EDMPI devices about a fixed center of rotation were developed by Aesculap Implant Systems. Their design was shared, and a set of these fixtures were manufactured and evaluated independently at Empirical Testing Corp. Testing was conducted to quantify the loads applied to an EDMPI in the FE and LB fixtures, determine whether results are frequency dependent, and evaluate whether EDMPI of different stiffnesses can be differentiated when using the fixtures. Miniature load cells were mounted to FE and LB fixtures to quantify load applied to an EDMPI during testing at several different applied loads and at two testing frequencies (2 Hz and 5 Hz). The load cells were replaced with cylindrical test specimens made of stainless steel, titanium, polyether ether ketone (PEEK), or a polyethylene cable slipped inside a polyurethane tube and the testing repeated. Results indicated that loads measured during FE testing were similar in magnitude to those recorded during LB testing but 90 degrees out of phase. No frequency effects were observed. Differences in displacement amplitude when testing the stainless steel, titanium, and PEEK specimens in FE and LB testing at both 2 Hz and 5 Hz were not discernable. The displacement amplitude of the polyethylene (PE) cord specimen was different than the other specimens, but no frequency effect was noted. The fixtures developed by Aesculap are a reasonable attempt to conduct EDMPI testing in FE, LB, and AR motions. Potential users of these fixtures should conduct further studies to determine whether their use is appropriate to test their specific EDMPI device.
机译:ASTM F2624-07规定了施加屈曲/伸展(FE),横向弯曲(LB)和轴向旋转(AR)力矩来测试和评估额外的椎间盘保留运动植入物(EDMPI),但是诸如夹具设计和载荷参数的参数保留了下来给用户定义。 Aesculap Implant Systems开发了一套夹具,用于进行EDMPI设备围绕固定旋转中心进行FE,LB和AR测试旋转。他们的设计是共享的,并且在Empirical Testing Corp.中独立制造和评估了一组这些夹具。进行了测试以量化FE和LB夹具中施加到EDMPI的负载,确定结果是否与频率相关,以及评估EDMPI使用固定装置时,可以区分不同的刚度。将微型称重传感器安装到FE和LB夹具上,以量化在测试期间在几种不同的施加载荷和两个测试频率(2 Hz和5 Hz)下施加到EDMPI的载荷。用不锈钢,钛,聚醚醚酮(PEEK)或滑入聚氨酯管的聚乙烯电缆制成的圆柱状试样代替称重传感器,然后重复测试。结果表明,有限元测试期间测得的负载与LB测试期间记录的负载大小相似,但相位差为90度。没有观察到频率影响。在2 Hz和5 Hz的FE和LB测试中测试不锈钢,钛和PEEK试样时,位移幅度的差异是无法分辨的。聚乙烯(PE)帘线样品的位移幅度与其他样品不同,但未观察到频率效应。 Aesculap开发的夹具是在FE,LB和AR运动中进行EDMPI测试的合理尝试。这些设备的潜在用户应进行进一步研究,以确定他们的使用是否适合测试其特定的EDMPI设备。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号