ASTM F2624-07 prescribes applying flexion/extension (FE), lateral bending (LB), and axial rotation (AR) moments to test and evaluate extra discal motion preserving implants (EDMPI), but parameters such as fixture design and load parameters are left to the user to define. A set of fixtures to conduct FE, LB, and AR testing rotation of EDMPI devices about a fixed center of rotation were developed by Aesculap Implant Systems. Their design was shared, and a set of these fixtures were manufactured and evaluated independently at Empirical Testing Corp. Testing was conducted to quantify the loads applied to an EDMPI in the FE and LB fixtures, determine whether results are frequency dependent, and evaluate whether EDMPI of different stiffnesses can be differentiated when using the fixtures. Miniature load cells were mounted to FE and LB fixtures to quantify load applied to an EDMPI during testing at several different applied loads and at two testing frequencies (2 Hz and 5 Hz). The load cells were replaced with cylindrical test specimens made of stainless steel, titanium, polyether ether ketone (PEEK), or a polyethylene cable slipped inside a polyurethane tube and the testing repeated. Results indicated that loads measured during FE testing were similar in magnitude to those recorded during LB testing but 90 degrees out of phase. No frequency effects were observed. Differences in displacement amplitude when testing the stainless steel, titanium, and PEEK specimens in FE and LB testing at both 2 Hz and 5 Hz were not discernable. The displacement amplitude of the polyethylene (PE) cord specimen was different than the other specimens, but no frequency effect was noted. The fixtures developed by Aesculap are a reasonable attempt to conduct EDMPI testing in FE, LB, and AR motions. Potential users of these fixtures should conduct further studies to determine whether their use is appropriate to test their specific EDMPI device.
机译:ASTM F2624-07规定施加屈曲/延伸(FE),横向弯曲(LB)和轴向旋转(AR)矩试验和评估额外的圆形运动保存植入物(EDMPI),但诸如夹具设计和负载参数等参数剩余向用户定义。通过Aesculap植入系统开发了一组传导Fe,LB和AR测试EDMPI器件的旋转的旋转旋转的旋转。它们的设计被共享,并在经验测试公司中独立制造和评估这些灯具。进行测试以量化应用于FE和LB夹具中的EDMPI的负载,确定结果是否依赖于频率,并评估EDMPI是否频率。在使用固定装置时可以区分不同的刚度。将微型称重传感器安装到Fe和LB夹具上,以在几种不同施加的载荷和两个测试频率(2Hz和5Hz)的测试期间量化施加到EDMPI的负荷。用由不锈钢,钛,聚醚醚酮(PEEK)制成的圆柱形试验样品代替称号,或者在聚氨酯管内滑动的聚乙烯电缆,重复测试。结果表明,在FE测试期间测量的负载的幅度相似于LB测试期间记录的那些,但相位的90度。没有观察到频率效应。在2Hz和5Hz两种Hz的Fe和LB测试中测试不锈钢,钛和PEEK标本时位移幅度的差异是不可辨别的。聚乙烯(PE)帘线标本的位移幅度不同于其他样本,但没有注意到频率效应。由Aesculap开发的固定装置是在Fe,LB和AR运动中进行EDMPI测试的合理尝试。这些固定装置的潜在用户应该进行进一步的研究,以确定它们是否适合测试其特定的EDMPI设备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号