An Assessment of the Performance and Quality Control Procedures of PACS Workstation Monitors used in Irish Radiology Departments

机译:对爱尔兰放射科使用的PACS工作站监视器的性能和质量控制程序的评估

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Diagnostic efficacy in soft-copy reporting relies heavily on the quality of workstation monitors and an investigation performed in 2002 demonstrated that CRT monitors in Dublin imaging departments were not operating at optimal levels. The current work examines the performance of CRTs being used in Dublin and other parts of Ireland to establish if problems reported in the earlier work have been rectified. All hospitals performing soft-copy reporting for general radiology using CRTs were included in the work. Examination of ambient lighting, calibration of monitors and analysis of CRT performance using the SMPTE test pattern and a selection of the AAPM test images was performed. Maximum luminance, spatial uniformity of luminance, temporal luminance stability, gamma, geometry, sharpness, veiling glare and spatial resolution of each monitor was evaluated. Ambient lighting in all reporting areas was within recommended levels. All the monitors were calibrated appropriately and were performing at acceptable levels for maximum luminance and temporal stability and only one of the thirty-three investigated failed to reach the standard for spatial uniformity. In contrast a number of the CRTs investigated showed poor adherence to acceptable levels for geometrical distortions, veiling glare and spatial resolution all of which are important influencers of image quality. Gamma values also appeared to be low for a number of monitors but this interpretation is provisional and subject to the establishment of ratified guideline values. The results demonstrate that although some improvement on the previous situation is evident, greater adherence to acceptable levels is required for certain parameters.
机译:软拷贝报告的诊断效力在很大程度上取决于工作站监视器的质量,2002年进行的一项调查表明,都柏林成像部门的CRT监视器未达到最佳水平。当前的工作检查了都柏林和爱尔兰其他地区使用的CRT的性能,以确定早期工作中报告的问题是否已得到纠正。所有使用CRT对普通放射学进行软拷贝报告的医院都包括在工作中。使用SMPTE测试图案和选择的AAPM测试图像进​​行了环境照明的检查,监视器的校准以及CRT性能的分析。评估了每个监视器的最大亮度,亮度空间均匀性,时间亮度稳定性,伽玛,几何形状,清晰度,遮盖眩光和空间分辨率。所有报告区域的环境照明均在建议水平之内。所有的监视器都经过适当的校准,并且在最大亮度和时间稳定性方面处于可接受的水平,并且被研究的三十三个监视器中只有一个未能达到空间均匀性标准。相比之下,许多被研究的CRT显示出对几何变形,遮盖眩光和空间分辨率的可接受水平的依从性差,这些都是影响图像质量的重要因素。对于许多监视器来说,伽玛值似乎也很低,但是这种解释是临时性的,并有待批准的准则值确定。结果表明,尽管明显改善了以前的情况,但是某些参数仍需要更大程度地遵守可接受的水平。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号